EDIT. DATE DESIG. DESCRIPTION APPRO.
FANUC LTD 006/10
15323 Calibration time for power calibration (This is enable at EGE=1)
Time of power command to get power calibration efficiency
Unit : Sec
Setting range : 0 - 32767
Standard setting : 180: C1000-E, C1500B, C2000B, C2000C, C2000-E
300: C3000C, C3000D, C4000A, C4000-E, C6000A,C6000B
15324 Tube pressure for power calibration(50Hz)(This is enable at EGE=1)
Setting value of gas pressure for power calibration
This is the same as PRM15325 for oscillator installed inverter.
(This is setting value for oscillator not installed inverter in the area of 50Hz.)
Unit : 0.1 Torr
Setting range : 0 – 32767
Standard setting : 570: C6000A,C6000B
600: C1000-E
660: C1500B,C2000C, C2000-E
690: C2000B
700: C3000C
730: C4000A, C4000-E
770: C3000D
15325 Tube pressure for power calibration(60Hz)(This is enable at EGE=1)
Setting value of gas pressure for power calibration
(This is setting value for oscillator installed and not installed inverter in the area of 60Hz)
Unit : 0.1Torr
Setting range : 0 – 32767
Standard setting : 570: C6000A,C6000B
660: C1500B,C2000C, C2000-E
690: C2000B
700: C3000C
730: C4000A, C4000-E
770: C3000D
Note) Please set values of PRM15242,
15243 into PRM15324, 15325, because gas
pressure is sometimes changed in the
Note) Please set values of PRM15242,
15243 into PRM15324, 15325, because gas
pressure is sometimes changed in the
TITLE
A-81053E-053
SHEET
AUTOMATIC AGING FUNCTION
DRAW. No.