B-70315EN/02 APPENDIX E.PARAMETER LIST
- 325 -
E.12 PARAMETERS FOR GAS CONTROL (1)
15240 Negative pressure in exhaust completion
[Data type] Word
[Unit of data] (1=13Pa)
[Valid data range] 0 to 32767
This pressure is used for evacuating the discharge tubes in laser
start-up sequence.
15241 Intra-tube pressure in discharge start
[Data type] Word
[Unit of data] (1=13Pa)
[Valid data range] 0 to 32767
Internal gas pressure of the discharge tubes at the start of discharge
15242 Tube pressure for oscillation (50Hz)
[Data type] Word
[Unit of data] (1=13Pa)
[Valid data range] 0 to 32767
Internal gas pressure of the discharge tubes during laser oscillation. (at
50 Hz)
15243 Tube pressure for oscillation (60Hz)
[Data type] Word
[Unit of data] (1=13Pa)
[Valid data range] 0 to 32767
Internal gas pressure of the discharge tubes during laser oscillation. (at
60 Hz)
15244 Gas pressure control gain
[Data type] Word
[Unit of data]
[Valid data range] 0 to 32767
This is the gain to feedback control the intra-tube laser gas pressure.
15245 Intratube pressure control gain 2
[Data type] Word
[Unit of data]
[Valid data range] 0 to 32767
Set the gas pressure control gain during base discharge.
When bit 1 of parameter No.15000 is 1, this parameter is valid.