Page 81. OVERVIEW B–70115EN/03 1 OVERVIEW This manual describes the maintenance of the FANUC LASER C series, as well as the structure, configuration, and operation of the laser oscillator. This manual is aimed at those personnel responsible for laser oscillator maintenance. 6
Page 9B–70115EN/03 1. OVERVIEW 1.1 This manual is organized as described below. ORGANIZATION OF 1. Overview This chapter describes the organization of this manual, applicable THE MANUAL models, related manuals, and notes on reading this manual. 2. Safety This chapter describes the handling of lasers, and
Page 11B–70115EN/03 1. OVERVIEW 1.3 The following manuals are available for the FANUC LASER C series : RELATED MANUALS FANUC Series 0–L DESCRIPTIONS B–61572E CONNECTION MANUAL B–61573E OPERATOR’S MANUAL B–61574E MAINTENANCE MANUAL B–61575E FANUC Series 16–LA DESCRIPTIONS B–61852E CONNECTION MANUAL B–61853E
Page 121. OVERVIEW B–70115EN/03 1.4 In this manual, the words Warning and Caution, corresponding to different levels of safety requirements, are provided to ensure the user’s TO USE THE LASER safety and protect against damage to the laser oscillator. OSCILLATOR Moreover, sections entitled Note are used to
Page 13B–70115EN/03 2. SAFETY 2 SAFETY C1500B, C2000C (C2000B), C3000D (C3000C), C4000A, C6000B produce the rated output laser power of 1500W, 2000W, 3000W, 4000W, 6000W. The CO2 laser beam is the wavelength of 10.6 mm, far infrared, and is invisible to human eyes. The adequate care must be taken when CO2
Page 142. SAFETY B–70115EN/03 2.1 1) Potential hazards Laser oscillator emits CO2 laser beam(10.6 mm), which is high power LASER BEAM and invisible. D Being directly exposed to the CO2 laser beam could severely burn you. D The CO2 laser beam could be scattered by your workpiece and burn your eyes or skin.
Page 15B–70115EN/03 2. SAFETY Fig. 2.1 (c) is the position of panel that laser beam exposure is occurred without panel in C6000B, when you maintain the oscillator. Fig. 2.1 (d) is the beam path in the cabinet of C1500B, C2000B, C2000C. Fig. 2.1 (e) is the beam path in the cabinet of C3000C, C3000D. Fig. 2.
Page 162. SAFETY B–70115EN/03 Fig. 2.1 (c) The position of panel that laser beam exposure is occurred without panel in C6000B. Fig. 2.1 (d) The beam path in the cabinet of C1500B, C2000B, C2000C. 14
Page 17B–70115EN/03 2. SAFETY Fig. 2.1 (e) The beam path in the cabinet of C3000C, C3000D. Fig. 2.1 (f) The beam path in the cabinet of C4000A. 15
Page 182. SAFETY B–70115EN/03 Fig. 2.1 (g) The beam path in the cabinet of C6000B. 16
Page 19B–70115EN/03 2. SAFETY 2.2 1) Potential hazards There is RF voltage of 3 to 4kVo–p in the cabinet of the laser oscillator. HIGH VOLTAGE There is 200 VAC power in the relay panel, be careful not to touch the high voltage. 2) Safety recommendations When it checks the oscillator and exchange the unit,
Page 202. SAFETY B–70115EN/03 Front side Back side Fig. 2.2. (a) The position of high voltage in C1500B (Front side, Back side) 18
Page 21B–70115EN/03 2. SAFETY Front side Back side Fig. 2.2. (b) The position of high voltage in C2000B (Front side, Back side) 19
Page 222. SAFETY B–70115EN/03 Front side Back side Fig. 2.2.(c) The position of high voltage in C2000C (Front side, Back side) 20
Page 23B–70115EN/03 2. SAFETY Front side Back side Auxiliary unit Fig. 2.2.(d) The position of high voltage in C3000C(Front , Back , AUX ) 21
Page 242. SAFETY B–70115EN/03 Front side Back side Fig. 2.2.(e) The position of high voltage in C3000D(Front side, Back side) 22
Page 25B–70115EN/03 2. SAFETY Front side Back side Fig. 2.2.(f) The position of high voltage in C4000A (Front side, Back side) 23
Page 262. SAFETY B–70115EN/03 Front side Back side Fig. 2.2.(g) The position of high voltage in C6000B (Front side, Back side) 24
Page 27B–70115EN/03 2. SAFETY 2.3 1) Potential hazards CO2 laser beam is delivery from oscillator. Direct or scattered beam SAFETY is exposed. ENCLOSURE (AT 2) Safety recommendations YOUR WORK Mount the safety enclosure made of acrylic resin which can absorb the STATION) laser beam around the working envir
Page 282. SAFETY B–70115EN/03 2.6 1) Potential hazards When you touch a part of high temperature, your skin burn. HIGH TEMPERATURE 2) Safety recommendations The pipes of the gas circular system are very a high temperature. Do not touch pipes, heat exchanger and turbo blower because it does not do the burn.
Page 29B–70115EN/03 2. SAFETY Front side Back side Fig. 2.6 (a) The position of high temperature in C1500B (Front side, Back side) 27
Page 302. SAFETY B–70115EN/03 Front side Back side Fig. 2.6 (b) The position of high temperature in C2000B (Front side, Back side) 28
Page 31B–70115EN/03 2. SAFETY Front side Back side Fig. 2.6 (c) The position of high temperature in C2000C (Front side, Back side) 29
Page 322. SAFETY B–70115EN/03 Front side Back side Auxiliary unit Fig. 2.6 (d) The position of high temperature in C3000C (Front , Back, Aux) 30
Page 33B–70115EN/03 2. SAFETY Front side Back side Fig. 2.6 (e) The position of high temperature in C3000D (Front side, Back side) 31
Page 342. SAFETY B–70115EN/03 Front side Back side Fig. 2.6 (f) The position of high temperature in C4000A (Front side, Back side) 32
Page 35B–70115EN/03 2. SAFETY Front side Back side Fig. 2.6 (g) The position of high temperature in C6000B (Front side, Back side) 33
Page 362. SAFETY B–70115EN/03 2.7 Fig. 2.7 (a)–(g) show the location of the warning labels indicating the high voltage and laser beam path. WARNING LABELS Fig. 2.7 (a) is the location of the warning sticker at front side. (C1500B, C2000B, C2000C, C3000C, C3000D) Fig. 2.7 (b) is the location of the warning
Page 37B–70115EN/03 2. SAFETY Fig. 2.7 (b) The location of warning sticker at back side (C1500B, C2000B, C2000C, C3000C, C3000D) Fig. 2.7 (c) The location of the warning sticker (Auxiliary unit of C3000C) 35
Page 382. SAFETY B–70115EN/03 Fig. 2.7 (d) The location of warning sticker at front side. (C4000A) Fig. 2.7 (e) The location of warning sticker at back side. (C4000A) 36
Page 39B–70115EN/03 2. SAFETY Fig. 2.7 (f) The location of warning sticker at front side. (C6000B) Fig. 2.7 (g) The location of the warning sticker at back side. (C6000B) 37
Page 402. SAFETY B–70115EN/03 D Detail of warning sticker 1) Warning logotype (C1500B) 7000W 38
Page 422. SAFETY B–70115EN/03 5) Caution label for lifting 6) Aperture label 7) Label of non–interlocked protective panel 40
Page 43B–70115EN/03 2. SAFETY 8) Identification label 9) Address label 10) High voltage warning label 41
Page 442. SAFETY B–70115EN/03 11) Supply voltage label 12) Label of over–current protective 42
Page 45B–70115EN/03 2. SAFETY 13) Label of motor and transformer (C1500B,C2000B,C2000C) 13) Label of motor and transformer (C3000C) 43
Page 462. SAFETY B–70115EN/03 13) Label of motor and transformer (C3000D) 13) Label of motor and transformer (C4000A) 44
Page 47B–70115EN/03 2. SAFETY 13) Label of motor and transformer (C6000B) 14) Label of warning light 15) Maintenance label 45
Page 482. SAFETY B–70115EN/03 16) Certification label 17) Short–circuit interrupting capacity of main breaker 18) Caution label for lifting 46
Page 49B–70115EN/03 2. SAFETY 2.8 All the laser products have to comply with the various kinds of laser safety regulations, which include the use of key control. For instance, FDA KEY CONTROL PART 1040 PERFORMANCE STANDARDS FOR LIGHT–EMITTING PRODUCTS, Sec 1040. 10 (f), (4) states: “Each laser system class
Page 502. SAFETY B–70115EN/03 2.10 Press the emergency stop button when it is dangerous and breaks down. The oscillator is stopped discharging, gas pressure control and stand by EMERGENCY STOP purge state. BUTTON Use the one with the compulsion dissociation mechanism for the relay used for the emergency st
Page 523. INTERNAL STRUCTURE B–70115EN/03 3.1 Fig. 3.1 (a) to (g) show the block diagram of the laser oscillator. The FANUC LASER C series consists of a laser resonator, laser excitation GENERAL power supply, forced gas circulating system, pressure controller, exhaust controller, CNC interface, and a prote
Page 57B–70115EN/03 INPUT UNIT STABILIZED PSU LASER LASER LASER PSU UNIT PSU UNIT PSU UNIT SIGNAL CABLE MATCHING MATCHING MATCHING MATCHING DEW BEAM (TO CNC) BOX BOX BOX BOX SENSOR FOLDING UNIT GUIDE FOLDING LASER OUTPUT MIRROR COUPLER DISCHARGE DISCHARGE TUBE FOLDING TUBE MIRROR SHUTTER 55 UNIT FOLDING DI
Page 58INPUT UNIT LASER LASER LASER LASER STABILIZED PSU UNIT PSU UNIT PSU UNIT PSU UNIT SIGNAL PSU DEW BEAM CABLE MATCHING MATCHING MATCHING MATCHING SENSOR FOLDING BOX BOX BOX BOX 3. INTERNAL STRUCTURE (TO CNC) UNIT GUIDE LASER OUTPUT COUPLER FOLDING MIRROR FOLDING SHUTTER MIRROR UNIT FOLDING MIRROR FOLD
Page 59B–70115EN/03 INPUT UNIT STABILIZED SIGNAL PSU CABLE (TO CNC) GUIDE FOLDING LASER MIRROR 57 FOLDING MIRROR POWER SENSOR FROM INLET INLET INLET INLET TO PARTS PARTS TURBO Fig. 3.1 (g) Block diagram (C6000B) H/E H/E PCB H/E H/E 3. INTERNAL STRUCTURE
Page 603. INTERNAL STRUCTURE B–70115EN/03 3.2 The following describes the details of each component of the FANUC LASER C series. Fig. 3.2 (a) to (p) show the internal structure. COMPONENT DETAILS (1) Resonator The resonator consists of an output mirror, rear mirror, folding mirrors, discharge tubes, power
Page 61B–70115EN/03 3. INTERNAL STRUCTURE (11) Pressure control unit This unit constantly monitors the gas pressure in the discharge tubes and supplies fresh laser gas to the gas circulating system, thus maintaining a constant pressure in the discharge tubes. This unit also monitors the laser gas supply st
Page 623. INTERNAL STRUCTURE B–70115EN/03 (19) Water distribution unit This unit distributes cooling water, supplied from either a chiller unit or a temperature–regulated external water supply, to each unit in the laser oscillator. For safety, the water distribution unit is equipped with a flow sensor whic
Page 63B–70115EN/03 3. INTERNAL STRUCTURE (29) Control PCB This PCB sends the contractor open/close signal to the power magnetics cabinet, as directed by commands received from the CNC. It also notifies the CNC of the open/close status of the circuit breaker in the power magnetics cabinet. (30) Dew sensor
Page 744. INSTALLATION B–70115EN/03 4.1 Use the following procedure to make adjustments and checks during installation. INSTALLATION (1) Check the environment at the installation location. PROCEDURE D Environmental conditions 1) Ambient temperature : +5°C to +30°C 2) Temperature variation : 1.1°C/minute ma
Page 79B–70115EN/03 4. INSTALLATION (3) Check the units installed in the main unit cabinet. D Check items 1) Check whether any printed circuit boards are loose or removed. 2) Check whether any cables are damaged (such as damaged sheathing). 3) Check whether any connectors are loose or detached. 4) Check th
Page 804. INSTALLATION B–70115EN/03 3) Laser gas composition and purity (See Section 4.4.4.2.) 4) Quality of cooling water (See Section 4.4.4.1.) 5) Flow of cooling water (IN, OUT) (9) Check the input power supply voltage, frequency and phase sequence. D Check items 1) 200/220 VAC +10%, –15%, 50/60 Hz "1 H
Page 81B–70115EN/03 4. INSTALLATION 4) Provided no water leakage is observed in step 3) above, allow cooling water to flow through the oscillator at the specified flow rate. Set the output pressure of the cooling water circulating unit to 0.5 MPa. Then, check that there is no water leakage at the following
Page 824. INSTALLATION B–70115EN/03 2) If the laser gas flow rate is abnormal, perform adjustment as described in Section 9.5. 3) Reset the values of the bits of parameter No. 15025 [201] through parameter No. 15028 [203] to their original values. This completes the check. (15) Perform laser oscillation to
Page 83B–70115EN/03 4. INSTALLATION 4.2 Prior to shipment and transportation, the packing and checking operations described below must be performed. PREPARATION PRIOR TO SHIPMENT (1) Disconnect all NC connecting cables. (2) Disconnect the power cable and ground cable of the input unit in the main unit cabi
Page 844. INSTALLATION B–70115EN/03 4.2.2 Remove cooling water according to the procedure below. Removing Cooling (1) Open the cooling water inlet (IN) and outlet (OUT), and leave both Water open. Cooling water will stop draining from the unit after about 10 minutes. (2) Once the cooling water has stopped
Page 85B–70115EN/03 4. INSTALLATION Fig 4.2.2 Detail of Distributor (C4000A) (5) Next, connect the compressed air hose to the cooling water outlet (OUT). At this time, ensure that cooling water inlet is open. Remove water in the same procedure as (2), (3), and (4). (6) Once all cooling water has been remov
Page 864. INSTALLATION B–70115EN/03 4.3 (1) Oscillation characteristics DETAILS OF 1) Check that bit 4 of parameter No. 15000 (16–L or 16i–L) or bit 4 of parameter No. 200 (0–L) is set to 1. CHECKING 2) Start the oscillator. When a discharge start switch is provided, do not set the switch to ON until after
Page 87B–70115EN/03 4. INSTALLATION 3) Remove the maintenance panel from the oscillator so that the discharge tubes can be checked. Set the door interlock to OFF. 4) Start the oscillator. When a discharge start switch is provided, do not set this switch to ON until after LRDY has been set. 5) Check that th
Page 884. INSTALLATION B–70115EN/03 6) Upon the completion of the check, restore the parameters modified in step 1) to their original values. An alarm, if issued in step 5), indicates an error. Check the external laser gas piping and laser gas composition, or perform aging. (4) Beam mode check 1) Remove th
Page 89B–70115EN/03 4. INSTALLATION (5) External gas piping leakage check (clamp test) 1) Using an approved gas pipe, connect the gas inlet of the oscillator to the cluster piping, or to the secondary side of the regulator mounted on the gas cylinder. CAUTION Rubber hoses were once commonly used for laser
Page 904. INSTALLATION B–70115EN/03 (6) Check for leakage within the oscillator D Negative pressure method 1) On the setting screen, enable parameter rewriting. 2) On the parameter screen, change the setting of parameter No. 15240 (16–L, 16i–L) or parameter No. 235 (0–L) to 0. 3) Set the oscillator start s
Page 91B–70115EN/03 4. INSTALLATION D Pressurization CAUTION The turbo blower is not designed to withstand pressurization. So, whenever possible, do not perform a leakage check by pressurizing the turbo blower. Even when internal pressurization is inevitable, never apply a pressure in excess of 0.08 MPa. O
Page 924. INSTALLATION B–70115EN/03 CAUTION 1 If the oscillator is started with an incorrect parameter value specified, there is a risk of the oscillator being damaged or destroyed. Be particularly careful when entering and checking parameter values. 2 Some parameters may require modification to, for examp
Page 93B–70115EN/03 4. INSTALLATION 2) Set the RUN key to ON to start the oscillator. 3) Perform discharge by setting the HV ON switch to ON. Check that discharge is performed continuously. 4) Then, perform internal discharge using the settings indicated below. Two methods of internal discharge are support
Page 944. INSTALLATION B–70115EN/03 4.4 The oscillator has connections for cooling water piping, laser gas piping, and power and signal cables. For details of the electrical connections for OSCILLATOR the NC and machine, refer to the corresponding connection manuals. CONNECTIONS FANUC Series 0–L CONNECTION
Page 95B–70115EN/03 4. INSTALLATION Fig. 4.4.1 (b) Water and laser gas connection (C2000C) 93
Page 964. INSTALLATION B–70115EN/03 Fig. 4.4.1 (c) Water and laser gas connection (C3000C) 94
Page 97B–70115EN/03 4. INSTALLATION Fig. 4.4.1 (d) Water and laser gas connection (C3000D) Fig. 4.4.1 (e) Water and laser gas connection (C4000A) 95
Page 984. INSTALLATION B–70115EN/03 Fig. 4.4.1 (f) Water and laser gas connection (C6000B) 4.4.1.1 The quality of cooling water is specified in the table below. If tap water Specification of the is used, it should be treated in an ion exchanger. cooling water Refrigerator/air–conditioner cooling water qual
Page 99B–70115EN/03 4. INSTALLATION 4.4.1.2 To avoid cooling water trouble and minimize the frequency of cooling Water treating agent water exchange, the following anticorrosive should be added to the cooling water. Consult the chiller manufacturer for use of the anticorrosive. Product name: CONTLIME K–600
Page 1004. INSTALLATION B–70115EN/03 4.4.1.4 If the chiller is used in a cold district, it should be provided with an Anti–freezing agent antifreezing function. When it is extremely cold, the chiller should be kept running. If it is necessary to use an antifreezing solution for lack of an alternative, the f
Page 101B–70115EN/03 4. INSTALLATION 4.4.2 Laser Gas 4.4.2.1 Supply the laser oscillator with a mixture of gases that satisfy the Laser gas specification conditions listed below. (1) Composition ratio and accuracy CO2 : 5 "0.25% He : 40 "2.00% N2 : 55 "2.75% (N2 balance) (2) Water (H2O): 5 ppm or less (3) H
Page 1024. INSTALLATION B–70115EN/03 4.4.3 The laser oscillator incorporates a safety circuit. The safety circuit is Electrical Connection enabled by using a terminal board (XT20). 4.4.3.1 Figs. 4.4.3.1 (a) to (g) show the cable inlet and the point of connection. NC–to–oscillator Attach the following cable
Page 103B–70115EN/03 4. INSTALLATION Fig. 4.4.3.1 (a) Cable connection of C1500B Fig. 4.4.3.1 (b) Cable connection of C2000B 101
Page 1044. INSTALLATION B–70115EN/03 Fig. 4.4.3.1 (c) Cable connection of C2000C Fig. 4.4.3.1 (d) Cable connection of C3000C 102
Page 105B–70115EN/03 4. INSTALLATION Fig. 4.4.3.1 (e) Cable connection of C3000D Fig. 4.4.3.1 (f) Cable connection of C4000A 103
Page 1064. INSTALLATION B–70115EN/03 Fig. 4.4.3.1 (g) Cable connection of C6000B c) OFF interlock signals (OFI1 and OFI2) Making the power supply for the oscillator ready to be switched off turns off these signals. Connect them in parallel with an NC OFF switch. Contact current–carrying capacity: 3 A on 250
Page 107B–70115EN/03 4. INSTALLATION b) SHL1 and SHL2 These terminals are for shutter interlocking. Strapping these terminals enables the shutter to be opened and closed. Contact current–carrying capacity: At least 1.2 A on 24 VDC Wire: 1.25 mm2, 25 m or shorter c) IB1 and IB2 These terminals are for semico
Page 1084. INSTALLATION B–70115EN/03 4.4.3.2 The figure shows the safety circuit for each oscillator. Details of safety circuit Fig. 4.4.3.2 (a) Safety circuit of C1500B 106
Page 115B–70115EN/03 4. INSTALLATION 4.4.4 The C3000C has an oscillator section and auxiliary machine section. Inter–unit Connections Interconnection is made according to the procedure described below. (C3000C Only) 4.4.4.1 (1) Power cable (CP200) Electrical cables This cable feeds 200 VAC to the high–frequ
Page 1164. INSTALLATION B–70115EN/03 4.4.4.4 (1) Connect gas OUT of the oscillator section to gas IN of the auxiliary Gas piping machine section, using the gas pipe provided with the oscillator. (2) Supply laser gas through gas IN of the oscillator section. Exhaust gas from the exhaust pump is output throug
Page 117B–70115EN/03 5. MAINTENANCE 5 MAINTENANCE In FANUC LASER C series, periodic inspection items have been reduced, and adjustments have been made easy. To keep the oscillator in a satisfactory operating condition over a long period, however, it is necessary to carry out periodic maintenance (including
Page 1185. MAINTENANCE B–70115EN/03 5.1 Table 5.1 lists daily inspection items. Inspect the FANUC LASER C series according to this table. When parts (including oil) have been used DAILY INSPECTION for a prescribed period, replace them quickly. Table 5.1 Daily inspection items for FANUC LASER C series Item P
Page 119B–70115EN/03 5. MAINTENANCE 5.2 The FANUC LASER C series contains consumables that must be replaced periodically. Table 5.2 (a), (b) lists such consumables and the PERIODIC related periodic maintenance work. Perform periodic maintenance as MAINTENANCE well as daily inspection described in Section 5.
Page 1205. MAINTENANCE B–70115EN/03 Table 5.2 (b) Periodic maintenance items and periods Interval of maintenance (Operation hour) Item C4000A C6000B 1 Output mirror cleaning 800 to 1200h 800 to 1200h 2 Rear mirror cleaning 3 Folding mirror cleaning 3000 to 4000h 1000h or processing quality go 4 0–shift mirr
Page 121B–70115EN/03 5. MAINTENANCE 5.3 When opening the panels and doors during maintenance, keep the power turned off. DETAILS OF Before replacing oil, be sure to check that purging is completed. MAINTENANCE Fig. 5.3 (a) Oil gauge of Turbo Blower and Exhaust pump of C1500B Fig. 5.3 (b) Oil gauge of Turbo
Page 1225. MAINTENANCE B–70115EN/03 Fig. 5.3 (c) Oil gauge of Turbo Blower and Exhaust pump of C2000C Fig. 5.3 (d) Oil gauge of Turbo Blower and Exhaust pump of C3000C 120
Page 123B–70115EN/03 5. MAINTENANCE Fig. 5.3 (e) Oil gauge of Turbo Blower and Exhaust pump of C3000D Fig. 5.3 (f) Oil gauge of Turbo Blower and Exhaust pump of C4000A 121
Page 1245. MAINTENANCE B–70115EN/03 Fig. 5.3 (g) Oil gauge of Turbo Blower and Exhaust pump of C6000B 5.3.1 (1) Check method Turbo Blower Oil Check the amount of oil in the turbo blower while referring to the figure below. The oil level should be between graduations H and L. This check should be made when t
Page 125B–70115EN/03 5. MAINTENANCE 3) Clean the oil inlet, hexagonal–head screw of the oil inlet, and O–ring by wiping with a clean cloth or paper. Ensure that these parts are completely free of dust. If the oil is contaminated with dust, the turbo blower may fail. Set the O–ring in the groove around the h
Page 1265. MAINTENANCE B–70115EN/03 5.3.2 (1) Check method Exhaust Pump Oil Watch the oil gauge, and check that the oil level is between graduations L and H. Also check whether the oil is dark. If the oil level is below L, add oil to the turbo blower or replace the oil in it. If the oil level is above H, dr
Page 127B–70115EN/03 5. MAINTENANCE 8) Put the maintenance panel back in place, and turn on the power. SCREW KNOB FILTER COVER O–RING SPRING WASHER EXHAUST FILTER ELEMENT OIL INLET PLUG Fig. 5.3.3 Exchange of exhaust pump filter 125
Page 1285. MAINTENANCE B–70115EN/03 5.3.4 Replace the filter every 1500 hours, or when oil mist occurs. Fig. 5.3.4 Exhaust System Filter shows the exhaust system filter. None of the C1500B, C2000B, C2000C, C3000C, C3000D, and C4000A is equipped with this filter. (1) Replacement method Clamp Housing head Hou
Page 129B–70115EN/03 5. MAINTENANCE 10) Take out a new cartridge filter (A97L–0201–0211, AdvanTech Toyo TCG–045–SIFS) from its bag, and put a plate gasket (supplied together with the cartridge filter) on both ends of the filter. 11) Insert the cartridge filter into the housing head, and fasten it with the f
Page 1305. MAINTENANCE B–70115EN/03 5.4 The maintenance parts are listed below. Refer to the following tables for maintenance unit and part specification information. MAINTENANCE PARTS NONREPAIRABLE PARTS MODEL QUANTITY NAME SPECIFICATION NOTE C1500B C2000B C2000C C3000C C3000D C4000A C6000B OUTPUT COUPLER
Page 131B–70115EN/03 5. MAINTENANCE MODEL QUANTITY NAME SPECIFICATION NOTE C1500B C2000B C2000C C3000C C3000D C4000A C6000B O RING FOR A98L–0003–0004/P22.4S 8 0 0 0 0 0 0 DISCHARGE TUBE A98L–0003–0004/P26S 0 8 8 12 12 16 0 A98L–0003–0004/P30.6S 0 0 0 0 0 0 24 O–RING FOR GAS A98L–0003–0004/P58 8 8 8 0 0 0 0
Page 137B–70115EN/03 5. MAINTENANCE MODEL QUANTITY NAME SPECIFICATION NOTE C1500B C2000B C2000C C3000C C3000D C4000A C6000B FET MODULE PCB A20B–1007–0140 1 A20B–1007–0150 1 This column is quantity of PSU parts per one LPSU. STEPUP A20B–1006–0072 1 (B210) TRANSFORMER PCB DIODE A50L–2001–0439 6 135
Page 141B–70115EN/03 5. MAINTENANCE #14 Use the service stock for the model specified in this list. Specify A20B–1005–0124 when newly ordering. #15 Use the service stock for the model specified in this list. Specify A91L–0001–0201 when newly ordering. When A91L–0001–0102 is installed, the following parts ar
Page 143B–70115EN/03 6. TROUBLESHOOTING 6.1 The troubleshooting procedure to be applied depends on the failure occurrence status as indicated below. Perform checking according to the TROUBLESHOOTING items listed below. PROCEDURE (1) Action in response to alarm screen display (See Sections 6.2 and 6.3.) (2)
Page 1446. TROUBLESHOOTING B–70115EN/03 6.3 The alarm number, DGN, and parameter number (PRM NO.) vary with the CNC model. So, the FS16–L or FS16i–L number comes first, and the RESPONDING TO FS0–L number is enclosed in parentheses. ALARM MESSAGES ON THE SCREEN Alarm No. 4061 [661] Anomaly of AD converter 1
Page 145B–70115EN/03 6. TROUBLESHOOTING Alarm No. 4063 [663] Anomaly of laser power supply unit (Refer to DGN 966, 967, 974, 975 [842, 854].) This alarm is issued, if a fault occurs in the laser power supply unit. The cause of the fault and the required action are described below. The laser power supply uni
Page 1466. TROUBLESHOOTING B–70115EN/03 No. Phenomenon Presumption cause 5 The RF voltage of all the units is high dur- The external laser gas ing discharge and the laser power is low. pipe is abnormal. The base discharge is narrow and tends to disappear. When the purge is repeatedly performed, the RF volta
Page 147B–70115EN/03 6. TROUBLESHOOTING Alarm No. 4065 [665] Shutter anomaly (Related DGNs are shown below.) The table below lists the valid combinations of shutter open and close commands and open and close position detection signals. The shutter position is detected three seconds after opening or closing
Page 1486. TROUBLESHOOTING B–70115EN/03 Alarm No. 4066 [666] Discharge anomaly (Refer to DGN = 909 to 914 [861 to 868] Pressing the HVON (start discharge) button causes the unit to start discharging. This alarm is issued, if at least one discharge tube fails to operate. No. Cause of trouble Solution 1 Anoma
Page 149B–70115EN/03 6. TROUBLESHOOTING Alarm No. 4067 [667] Overheat of laser cabinet (Refer to DGN = 961/bit 1 [841/bit7]) The temperature of the input side of the electrode cooling fan is monitored and sends alarm when it exceeds 60°C. When the temperature decreases, the alarm state is solved automatical
Page 1506. TROUBLESHOOTING B–70115EN/03 Alarm No. 4068 [668] Too much incident laser beam back to the resonator. (Refer to DGN = 906 [878]) This alarm is issued, if a workpiece reflects laser beam more than the rating to the laser oscillator. This can happen when the laser beam is used to drill, cut, or wel
Page 151B–70115EN/03 6. TROUBLESHOOTING Alarm No. 4071 [671] Anomaly of assist gas (Refer to DGN = G221/bit 7 [145/bit7] When starting machining, the NC monitors for a ready signal from the assist gas supply unit in the machine. If this signal is not sent normally, the NC issues this alarm. Check the DGN da
Page 1526. TROUBLESHOOTING B–70115EN/03 Alarm No. 4073 [673] Decrease of laser gas pressure (Refer to DGN = 960/bit 5 [840/bit 4]) The pressure of the laser gas supplied to the laser oscillator is monitored. This alarm is issued, if this pressure becomes lower than the permissible level (0.1 MPa). No. Cause
Page 153B–70115EN/03 6. TROUBLESHOOTING Alarm No. 4075 [675] Condensation (Refer to DGN = 961/bit 2 [840/bit 6]) There is a condensation sensor at the inlet of the water branch unit in the oscillator. This alarm is issued, when condensation is detected. No. Cause of trouble Solution 1 Tool low cooling water
Page 1546. TROUBLESHOOTING B–70115EN/03 Alarm No. 4077 [677] Overheat of beam absorber (Refer to DGN = 961/bit 7 [841/bit 3]) Laser beam is introduced into the beam absorber, when the oscillation takes place with shutter closed. This absorber is water–cooled and sends alarm when the temperature exceeds a cr
Page 155B–70115EN/03 6. TROUBLESHOOTING Alarm No. 4078 [678] Gas pressure anomaly in discharge tube (Refer to DGN = 905 [860]) The gas pressure is monitored after the discharge start ready condition (LRDY) is established. This alarm is issued, if the monitored gas pressure deviates by "100 (="1330 Pa) from
Page 1566. TROUBLESHOOTING B–70115EN/03 Alarm No. 4080 [680] Leakage of gas tube/anomaly of exhaust pump (Refer to DGN = 905 [860]) When the RUN (pressure control start) button is pressed, the exhaust pump starts operating and expels gas from the gas circulating system to cause a low–pressure condition. Thi
Page 157B–70115EN/03 6. TROUBLESHOOTING Alarm No. 4081 [681] Anomaly of gas pressure control (Refer to DGN = 905 [860] Pressing the RUN (pressure control start) button starts expelling air from the laser gas tube. When a specified degree of vacuum is attained, laser gas is supplied into the laser gas tube,
Page 1586. TROUBLESHOOTING B–70115EN/03 Alarm No. 4082 [682] Anomaly of pressure sensor (Refer to DGN = 905 [860] This alarm appears when the signal of the pressure sensor used for pressure control detector vanishes. No. Cause of trouble Solution 1 Disconnected cable be- Check the cable connection. If the c
Page 159B–70115EN/03 6. TROUBLESHOOTING Alarm No. – – – [684] Anomalous selection of the number of discharge tubes. Anomalous selection of the number of discharge tubes. (Only FS0–L) No. Cause of trouble Solution 1 The case of FS0–L. The number of discharge tubes should be The number of discharge selected t
Page 1606. TROUBLESHOOTING B–70115EN/03 Alarm No. 4087 [687] Shutter temperature abnormal (Refer to bit 6 of DGN 961 [bit 2 of 841].) This alarm is issued when the temperature of the shutter mirror exceeds a preset maximum. The table below lists the causes of faults and the corresponding countermeasures. No
Page 161B–70115EN/03 6. TROUBLESHOOTING Alarm No. 4090 [690] Laser beam not generated (Refer to bit 6 of DGN F221 [bit 6 of 172].) This alarm is issued, if an attempt is made to radiate a laser beam, when the laser oscillator is not in the LSTR state (discharging state). Alarm No. 4094 [– – –] Exhaust pump
Page 1626. TROUBLESHOOTING B–70115EN/03 Alarm No. 4100 [691] Inverter abnormal If the inverter used to power the turbo blower is abnormal, this alarm is issued. When this alarm is issued, check that the alarm LED on the LED indicator section of the inverter is lit. At this time, do not turn off the power to
Page 163B–70115EN/03 6. TROUBLESHOOTING Alarm No. Turbo blower not stopped 4103, 4113 [691] This alarm is issued if a turbo blower is not stopped within 120 seconds after RUN OFF. The power can be turned off even if the purge complete signal is not turned on. No. Cause of trouble Solution 1 Cable connection
Page 1646. TROUBLESHOOTING B–70115EN/03 Alarm No. 4107 [621] External reflecting mirror not installed This alarm is issued, if the beam reflecting unit is not attached with a mirror or mirror holder. No. Cause of trouble Solution 1 Mirror not mounted Mount a 0–degree shift mirror. 2 Sensor abnormal Replace
Page 165B–70115EN/03 6. TROUBLESHOOTING 6.4 MAJOR FAULTS 6.4.1 (1) If an alarm is issued to indicate the occurrence of a power supply fault, Laser Power Supply check the diagnosis screen to determine which power supply unit issued that alarm signal. The bit corresponding to the power supply Alarm Display un
Page 1666. TROUBLESHOOTING B–70115EN/03 (4) Table 6.4.1 explains each type of alarm. Table 6.4.1 Causes of power supply alarms Type of alarm Source Major causes DCI 1 This alarm is issued if the DC Load short–circuit DCI 2 current of the power supply unit DCI 3 is high when the oscillator is op- Discharge t
Page 167B–70115EN/03 6. TROUBLESHOOTING (5) If an alarm is issued at restart or at the start of discharge (HV ON), the power supply unit has been damaged. 1) When only the DCI alarm is issued The high–speed diode or FET module may be damaged. Locate the problem and replace any faulty component or module. 2)
Page 1686. TROUBLESHOOTING B–70115EN/03 6.4.5 First check if PRM No. 15216 [262] is taking the value of 60, which is Display of Fluctuating a standard. In case the fluctuation continues with this value used, read the laser output monitor value indicated in DGN=906 [878]. Its range of Laser Output on CRT flu
Page 169B–70115EN/03 6. TROUBLESHOOTING 6.4.7 Electromagnetic No. Cause of trouble Solution Contactor of Vane 1 Anomalous setting Ascertain if the setting of the contacter is done at Pump Trips Thermally. of thermal relay 5.5 A. If not so, correct it. 2 Too high viscosity (1) Check if the room temperature i
Page 1706. TROUBLESHOOTING B–70115EN/03 6.4.9 Excessive Laser Gas No. Cause of trouble Solution Consumption 1 Gas leakage be- Ascertain the following. tween gas cylinder (1) Anomalous mounting of regulator. and oscillator (2) Anomalous connection between tube and reg- ulator. (3) Anomalous connection of gas
Page 171B–70115EN/03 6. TROUBLESHOOTING OC : Momentary overcurrent No. Cause of trouble Solution 1 Turbo blower or cable short–cir- Replace the short–circuited unit or cuited cable. 2 Turbo blower or cable short–cir- Replace the short–circuited unit or cuited with ground cable. 3 Invalid setting of inverter
Page 1726. TROUBLESHOOTING B–70115EN/03 (2) Inverter (A90L–0001–0464/CE, 0465/CE: Model JH300) When an alarm is issued, the digital panel displays a message indicating the type of alarm. The table below lists the messages, descriptions, and causes. Message Description Cause E01 Overcurrent during turbo blow
Page 173B–70115EN/03 6. TROUBLESHOOTING 6.5 OBSERVING VOLTAGE OF POWER LINE 6.5.1 Ascertain the followings, Measurement of 1) line voltage: 200/220 VAC, +10%, –15% Voltage 2) frequency: 50/60 Hz – 1 Hz 3) 3 phases Here the combination of 220 V and 50 Hz is prohibited. 6.5.2 Using a phase rotation sensor, L1
Page 1746. TROUBLESHOOTING B–70115EN/03 6.5.3 Ascertain if the voltage of each DC power unit is within rated range. Here Measurement of refer to Fig. 6.5.3 (a), (b), (c) for the external view of the IF PCB in the oscillator. Refer to Fig. 6.5.3 (d) for the external view of the stabilized Voltage of DC Power
Page 175B–70115EN/03 6. TROUBLESHOOTING (4) Stabilized power supply unit Fig. 6.5.3 (d) shows the external appearance of +24V stabilized power unit. Fuse of the stabilized power supply unit Designation of power supply unit A20B–1005–0120 Designation of fuse FU : A60L–0001–0175#5.0 (5.0A) FU : A60L–0001–0175
Page 1786. TROUBLESHOOTING B–70115EN/03 6.5.4 By using the IF PCB pins, check that each signal of the IF PCB is normal. Checking the IF PCB Fig. 6.5.3 (a) and Fig. 6.5.3 (b) show the pin locations. Signals FS0–L FS16–L Signal function FS16i–L COR COR Voltage to operate the laser power supply 0 V to 5V GPC G
Page 179B–70115EN/03 6. TROUBLESHOOTING 6.6 When the machine breaks down, it is necessary to know if the trouble is in the CNC, PMC, machine or laser. INDICATION OF Even when there does not exist machine failure, the machine might STATE BY MEANS OF hang–up for waiting of external signals. Then the check of
Page 1806. TROUBLESHOOTING B–70115EN/03 #7 #6 #5 #4 #3 #2 #1 #0 G223 POV7 POV6 POV5 POV4 POV3 POV2 POV1 POV0 [G147] POV7 – POV6 Power override signal (input) [CNC→PMC status display] #7 #6 #5 #4 #3 #2 #1 #0 F220 MWRM SHTONL SHTOFL [F171] MWRM Laser power alarm signal (output) SHTONL Shutter on state signal
Page 181B–70115EN/03 6. TROUBLESHOOTING 6.6.2 Laser Oscillator Status Display (FS16–L, 16i–L) 900 LASER POWER COMMAND DOIML0A Displays a value specified in a laser output power command and sent to the DA converter. 901 LASER BIAS COMMAND DOIML08 Displays a value specified in a preparatory discharge command
Page 1826. TROUBLESHOOTING B–70115EN/03 909 RF VOLTAGE 1 910 RF CURRENT 1 911 RF VOLTAGE 2 912 RF CURRENT 2 913 RF VOLTAGE 3 914 RF CURRENT 3 915 RF VOLTAGE 4 916 RF CURRENT 4 917 RF VOLTAGE 1 918 RF CURRENT 1 919 RF VOLTAGE 2 920 RF CURRENT 2 921 RF VOLTAGE 3 922 RF CURRENT 3 923 RF VOLTAGE 4 924 RF CURREN
Page 183B–70115EN/03 6. TROUBLESHOOTING FRQ Power frequency detection 0 : 50 Hz 1 : 60 Hz *AP Air pressure sensor 0 : Lower than air pressure 1 : Same as air pressure *MGP Laser gas pressure 0 : Low 1 : Normal *RBT Blower temperature 0 : Abnormal 1 : Normal *SFI Safety interlock 0 : Interlocked 1 : Normal #
Page 1846. TROUBLESHOOTING B–70115EN/03 #7 #6 #5 #4 #3 #2 #1 #0 962 *OH1 *TCA1 *M–R *L–V *AR1 *IAL1 DIIML02 *IAL1 Inverter alarm 1 0 : Abnormal 1 : Normal *AR1 Frequency signal 1 0 : Mismatch 1 : Match *L–V Turbo blower oil 0 : Oil level too low 1 : Normal *M–R External folding mirror 0 : Mirror not mounted
Page 185B–70115EN/03 6. TROUBLESHOOTING #7 #6 #5 #4 #3 #2 #1 #0 964 *OH3 *TCA3 *REV3 *VIB3 *AR3 *IAL3 DIIML04 *IAL3 Inverter alarm 3 0 : Abnormal 1 : Normal *AR3 Frequency signal 3 0 : Mismatch 1 : Match *VIB3 Vibration sensor 3 0 : Abnormal 1 : Normal *REV3 Rotation sensor 3 0 : Less than 8000 rpm 1 : 8000
Page 187B–70115EN/03 6. TROUBLESHOOTING IB Semiconductor laser–on command 0: OFF 1: ON FW Inverter start signal 0: Stopped 1: Started PCS Gas pressure PWM command LSCST Low speed A/D conversion start 0: OFF 1: ON #7 #6 #5 #4 #3 #2 #1 #0 974 PSAR8 PSAR7 PSAR6 PSAR5 PSAR4 PSAR3 PSAR2 PSAR1 R@LSDIC2 Power unit
Page 1886. TROUBLESHOOTING B–70115EN/03 6.6.3 Laser Oscillator Status Display (FS0–L) DGN No #7 #6 #5 #4 #3 #2 #1 #0 88 Output command value + Bias command 89 (CNC internal DO) Displays a value specified in a laser output power command and sent to the DA converter. DGN No #7 #6 #5 #4 #3 #2 #1 #0 90 Bias com
Page 189B–70115EN/03 6. TROUBLESHOOTING MVW Quantity of cooling water 0 : Abnormal 1 : Normal AP Atmospheric pressure sensor 0 : Less than standard value 1 : Greater than standard value DGN No #7 #6 #5 #4 #3 #2 #1 #0 841 CAT SHOF SHON DL1 ABT SHT IAL ENB (CNC internal DI) CAT Cabinet temperature 0 : Abnorma
Page 1906. TROUBLESHOOTING B–70115EN/03 DGN No #7 #6 #5 #4 #3 #2 #1 #0 846 ADDT2 847 EOC2 (CNC internal DI) ADDT2 Data read from the AD converter 2 EOC2 Conversion completion signal of the AD converter 2 0: Under conversion 1: Completion DGN No #7 #6 #5 #4 #3 #2 #1 #0 848 PCBID (CNC internal DI) PCBID PCB A
Page 191B–70115EN/03 6. TROUBLESHOOTING SEL4 – SEL0 AD converter 2 switching signal SEL4 SEL3 SEL2 SEL1 SEL0 Pressure in side tube 0 0 0 0 0 Output voltage of RF power supply unit (1) 0 0 0 0 1 Output current of RF power supply unit (1) 0 0 0 1 0 Output voltage of RF power supply unit (2) 0 0 0 1 1 Output c
Page 1926. TROUBLESHOOTING B–70115EN/03 GRDY Laser gas supply valve open/close signal 0 : Laser gas supply valve close 1 : Laser gas supply valve open VEN Exhaust valve open/close signal 0 : Exhaust valve close 1 : Exhaust valve open DGN No #7 #6 #5 #4 #3 #2 #1 #0 857 PSS8 PSS7 PSS6 PSS5 PSS4 PSS3 PSS2 PSS1
Page 193B–70115EN/03 6. TROUBLESHOOTING DGN No 877 EIN (CNC internal DI) EIN Input from tracing sensor. (Value which is not multiplied by the correction factor) +10V . . . . . . 2047 0V . . . . . . 0 –10V . . . . . . –2047 DGN No 878 LSADC (CNC internal DI) LSADC Power sensor output. (Unit W) (Value which i
Page 2107. OSCILLATOR CONNECTIONS B–70115EN/03 7.3 Fig. 7.3 (a) shows the flow of supply gas and exhaust gas in the laser oscillator. Figs. 7.3 (b) to (h) are system diagrams showing the piping in GAS PIPING the laser oscillator. Pressure value 100 (=1330Pa) or more Pressure value 300 to 700 (=4000 to9300Pa
Page 2188. UNIT CONFIGURATION B–70115EN/03 8 UNIT CONFIGURATION 216
Page 219B–70115EN/03 8. UNIT CONFIGURATION 8.1 The input unit consists of a power magnetics cabinet, interface PCB, and stabilized power supply. The power magnetics cabinet is configured as INPUT UNIT described below: (1) C1500B Input unit (C1500B) QF1 QF2 Thermal KM1 PSU Relay No. 1 Thermal Relay Thermal P
Page 2208. UNIT CONFIGURATION B–70115EN/03 Fig. 8.1 (a) Input unit (C1500B) 218
Page 2228. UNIT CONFIGURATION B–70115EN/03 Fig. 8.1 (b) Input unit (C2000B) 220
Page 223B–70115EN/03 8. UNIT CONFIGURATION (3) C2000C Input unit (C2000C) QF1 QF2 KM1 PSU No. 1 Thermal Relay PSU QF3 KM2 No. 2 QF4 KM3 Inverter Turbo blower QF5 KM4 AC FAN Thermal Exhaust Hour QF6 KM5 Relay pump meter Transformer Magnetic valve QF7 QF8 Stabilized PSU unit IF PCB CNC INPUT UNIT AS ALARM SHO
Page 2248. UNIT CONFIGURATION B–70115EN/03 Fig. 8.1 (c) Input unit (C2000C) 222
Page 225B–70115EN/03 8. UNIT CONFIGURATION (4) C3000C Input unit (C3000C) QF1 Thermal PSU QF2 KM1 Relay No. 1 Thermal relay Thermal PSU QF3 KM2 Relay No. 2 Thermal PSU QF4 KM3 Relay No. 3 Thermal PSU QF5 KM4 Relay No. 4 Thermal QF6 AUXLIARY UNIT Relay QF7 KM5 AC FAN Magnetic valve QF8 Transformer QF9 Stabil
Page 2268. UNIT CONFIGURATION B–70115EN/03 Fig. 8.1 (d) Input unit (C3000C) 1/2 224
Page 227B–70115EN/03 8. UNIT CONFIGURATION Fig. 8.1 (d) Input unit (C3000C AUX) 2/2 225
Page 2288. UNIT CONFIGURATION B–70115EN/03 (5) C3000D Input unit (C3000D) QF1 QF2 KM1 PSU No. 1 Thermal Relay PSU QF3 KM2 No. 2 KM3 PSU QF4 No. 3 Inverter Turbo QF5 KM4 blower Thermal Exhaust Hour QF6 KM5 Relay pump meter QF8 Intermediate unit Magnetic valve Transformer QF9 Stabilized PSU IF PCB QF7 KM6 AC
Page 229B–70115EN/03 8. UNIT CONFIGURATION Fig. 8.1 (e) Input unit (C3000D) 227
Page 2308. UNIT CONFIGURATION B–70115EN/03 (6) C4000A Input unit (C4000A) QF1 PSU QF2 KM1 No. 1 Thermal Relay PSU QF3 KM2 No. 2 KM3 PSU QF4 No. 3 PSU QF5 KM4 Inverter No. 4 KM5 Inverter Turbo QF6 blower Thermal Exhaust Hour QF7 KM6 Relay pump meter QF9 Intermediate unit Magnetic valve Transformer QF10 Stabi
Page 231B–70115EN/03 8. UNIT CONFIGURATION Fig. 8.1 (f) Input unit (C4000A) 229
Page 2328. UNIT CONFIGURATION B–70115EN/03 (7) C6000B Input unit (C6000B) QF1 PSU QF2 KM1 No. 1 Thermal Relay QF3 KM2 PSU No. 2 KM3 PSU QF4 No. 3 PSU QF5 KM4 No. 4 QF6 KM5 AC FAN Thermal Exhaust Hour QF7 KM6 Relay pump meter KM7 PSU QF8 No. 5 PSU QF9 KM8 No. 6 PSU QF10 KM9 No. 7 PSU QF11 KM10 No. 8 Turbo In
Page 233B–70115EN/03 8. UNIT CONFIGURATION Fig. 8.1 (g) Input unit (C6000B) 1/3 231
Page 2348. UNIT CONFIGURATION B–70115EN/03 Fig. 8.1 (g) Input unit (C6000B) 2/3 232
Page 235B–70115EN/03 8. UNIT CONFIGURATION Fig. 8.1 (g) Input unit (C6000B) 3/3 233
Page 2368. UNIT CONFIGURATION B–70115EN/03 8.2 Intermediate PCB B is used to group signals sent from the shutter section and to output the grouped signals to the interface PCB. CN59 is an output INTERMEDIATE terminal for connecting a measuring instrument used in alignment. PCB B Fig. 8.2 shows the layout of
Page 237B–70115EN/03 8. UNIT CONFIGURATION 8.3 Fig. 8.3 (a) shows installation, and Fig. 8.3 (b) shows piping of the pressure controller respectively. The functions of each of the components PRESSURE shown in the figures are outlined below: CONTROL UNIT (1) Pressure control valve This valve is provided to m
Page 2388. UNIT CONFIGURATION B–70115EN/03 Fig. 8.3 (a) Pressure control unit layout 236
Page 239B–70115EN/03 8. UNIT CONFIGURATION Fig. 8.3 (b) Block diagram of gas pressure control unit 237
Page 2409. SETTING AND ADJUSTMENT B–70115EN/03 9 SETTING AND ADJUSTMENT 238
Page 241B–70115EN/03 9. SETTING AND ADJUSTMENT 9.1 This unit is precisely adjusted at the factory and should not require any further adjustment. If the power supply unit, matching box, or discharge LASER POWER tube is replaced, however, the unit should be set and adjusted as described SUPPLY UNIT below. Whe
Page 2429. SETTING AND ADJUSTMENT B–70115EN/03 Enhanced pulse setting disable Parameter No. 15213: 32000 4) C2000C Output compensation disable Bit 4 of parameter No. 15000, parameter No. 15201 (bit 4 of parameter No. 200, parameter No. 231, No. 232): 0 Laser power feedback disable Parameter No. 15208 (param
Page 243B–70115EN/03 9. SETTING AND ADJUSTMENT 7) C6000B Output compensation disable Bit 4 of parameter No. 15000, parameter No. 15201 (bit 4 of parameter No. 200, parameter No. 231, No. 232): 0 Laser power feedback disable Parameter No. 15208 (parameter No. 233): 0 Maximum output command setting Parameter
Page 2449. SETTING AND ADJUSTMENT B–70115EN/03 9.1.2.3 (1) Start the oscillator, and wait 10 minutes in the LRDY state. Then set C3000C (step 1) HV ON and wait until the gas pressure is stabilized in the base discharge state. (Wait about 3 minutes. See DGN 905 [16–L, 16i–L] or DGN 860 [0–L].) Check that all
Page 245B–70115EN/03 9. SETTING AND ADJUSTMENT ÑÑÑ ÌÌÌÌÌÌÌ ÏÏ ÎÎÎÎ ÎÎÎÎ ÏÏÏ ÌÌÌÌÌÌ ÓÓÓ ÌÌÌÌÌÌÌ ÏÏ ÎÎÎÎ ÎÎÎÎ ÏÏÏ ÌÌÌÌÌÌ ÓÓÓ ÌÌÌÌÌÌÌ ÌÌÌÌÌÌ Power supplies 1, 2, 3, Power supplies 4, 5, 6, 7, 8, and 9 10, 11, and 12 Fig. 9.1.2.5 Base discharge adjustment points (indicated by arrows) (3) Check that the laser ou
Page 2469. SETTING AND ADJUSTMENT B–70115EN/03 9.1.2.6 (1) Start the oscillator, and wait until the gas pressure is stabilized in the C1500B (with a trigger base discharge state. (Wait about 3 minutes. See DGN 905 [16–L, 16i–L] or DGN 860 [0–L].) Then, check that all discharge tubes are electrode), C2000C,
Page 247B–70115EN/03 9. SETTING AND ADJUSTMENT Fig. 9.1.2.7 (a) Laser power supply unit layout (6 kW PSU) Fig. 9.1.2.7 (b) Adjust trimmer position (6 kW PSU) 245
Page 2489. SETTING AND ADJUSTMENT B–70115EN/03 Fig. 9.1.2.7 (c) Adjust trimmer position (9 kW PSU) 9.1.3 (1) Specify the output of a half of the rated output value, and turn on the beam. While checking that the power supply output (Note 4) does not Maximum Output exceed the maximum power supply output (Note
Page 249B–70115EN/03 9. SETTING AND ADJUSTMENT 9.1.5 (1) With the C2000B and C2000C, specify a pulse test command in the Pulse Output Test table of parameter No. 15207 and No. 15210. (2) Output the pulse test command specified in the table by 5 Hz and 5% for a pulse test time, and check that no alarm is iss
Page 2509. SETTING AND ADJUSTMENT B–70115EN/03 NOTE 4 Settings for each oscillator C1500B C2000B C3000C C3000C C6000B C6000B without a trigger (No.1, 4 PSU) (No.2, 3 PSU) (No. 1, 4, 5, 8 No. 2, 3, 6, 7 electrode PSU) PSU) S2 setting B201 B202 B201 B202 B202 B201 S5 setting SHORT SHORT SHORT SHORT USUAL SHOR
Page 251B–70115EN/03 9. SETTING AND ADJUSTMENT 9.1.7 Alarm level setting value of laser PSU is shown the following table. Alarm Level of Laser PSU Table 9.1.7 (a) Alarm level setting value Specification of laser PSU B204 B205 B206 B207 B208 B209 B210 DCI1 DC peak over 100A 100A (100A) 120A 120A 120A 120A cu
Page 253B–70115EN/03 9. SETTING AND ADJUSTMENT 9.2 These PCBs are precisely adjusted at the factory and should not require any further adjustment. If the turbo blower or turbo PCB is replaced, TURBO PCB however, VR1 to VR4 should be adjusted and set oil level sensor as described below. Turbo PCB: A16B–1600–
Page 2549. SETTING AND ADJUSTMENT B–70115EN/03 Adjustment position C1500B : VR1 only C2000B : VR1 only C2000C : VR1 only C3000C : VR1 VR2 only C3000D : VR1 only C4000A : VR1 only C6000B : VR1 to VR4 Fig. 9.2 (a) Detail of turbo PCB (A16B–1600–0670) 252
Page 2569. SETTING AND ADJUSTMENT B–70115EN/03 9.3 The inverter is precisely adjusted at the factory and should not require any further adjustment. If it is replaced, however, the high–frequency inverter HIGH–FREQUENCY should be set and adjusted as described below. INVERTER The type of the high–frequency in
Page 257B–70115EN/03 9. SETTING AND ADJUSTMENT (7) Adjust the acceleration/deceleration period. Adjust the period from when the turbo blower starts rotating until the rotation stabilizes, and the period from when deceleration starts until the turbo blower stops. Setting the acceleration 1) While the inverte
Page 2589. SETTING AND ADJUSTMENT B–70115EN/03 Selection of T1 or T10 of the DS1 switch determines the factor of the acceleration or deceleration time. When T10 is selected, the time is multiplied by 10. TX1: 3 to 15 sec TX10: 15 to 150 sec. DR ON DS2 OFF F.OS F60 F.FIN ON DB DS1 ACCEL DECEL T.BST F.STP D.B
Page 259B–70115EN/03 9. SETTING AND ADJUSTMENT 9.3.2 (1) Connection Adjusting the Inverter The following cables are connected to the inverter: R. S. T: Three–phase power supply (A90L–0001–0464/CE U. V. W: Turbo blower power supply and 0465/CE: Model H, O, L: Frequency command Name: JH300) 1: Reset signal 2:
Page 2609. SETTING AND ADJUSTMENT B–70115EN/03 D Frequency setting, multi–speed setting or other frequency displays, motor rotational speed display, frequency conversion display, or output current display. NOTE When conflicting data is set, a warning WARN.... will be displayed. For 6 seconds thereafter, do
Page 261B–70115EN/03 9. SETTING AND ADJUSTMENT D MONITOR/FUNCTION The operator has two modes; MONITOR mode and FUNCTION mode. mode In the MONITOR mode, the monitor screen to display operation status appears. In the function mode, the function screen appears. The function screen has a hierarchical structure
Page 2629. SETTING AND ADJUSTMENT B–70115EN/03 Frequency command Displays the current frequency command status. Operation mode Display Remote operator Terminal Optional PCB 1 Optional PCB 2 Monitor/function mode FR FT F1 F2 D Monitor screen display “MONITOR” key ST Monitor 29 OR FR " FS 0.00 Hz 0.00Hz S
Page 263B–70115EN/03 9. SETTING AND ADJUSTMENT D Function screen display “FUNCTION” key ST Monitor 29 OR FR " 1 Command " 2 Initial " Function mode menu screen 3 Function " (Function mode 1st layer) 4 Option " Next layer mark This mark indicates that there is one more layer to display. To go to the next
Page 2649. SETTING AND ADJUSTMENT B–70115EN/03 (3) Parameter settings The following table lists the parameters for which the values have been changed from the initial values. If the inverter has been shipped alone from FANUC, the user must change the parameters as listed below. To change the parameters, an
Page 265B–70115EN/03 9. SETTING AND ADJUSTMENT 415 "1 440 220 "1 250 Monitor (LED display) POWER Lamp This display shows frequen- Power lamp of cy, motor current, motor revo- control circuit lution speed, and trip history Up key, Down key These keys are used to change data and increase or decrease the frequ
Page 2669. SETTING AND ADJUSTMENT B–70115EN/03 9.4 This unit is precisely adjusted at the factory and should not require any further adjustment. If the pressure controller is replaced or requires PRESSURE re–adjustment, however, it should be set and adjusted as described below: CONTROL UNIT 9.4.1 (1) Turn t
Page 267B–70115EN/03 9. SETTING AND ADJUSTMENT PRESSURE ATMOSPHERIC SENSOR PRESSURE SWITCH LASER GAS PRESSURE SWITCH Fig. 9.4.2 Pressure control unit sensor layout 265
Page 2689. SETTING AND ADJUSTMENT B–70115EN/03 9.5 This unit is precisely adjusted at the factory and should not require any further adjustment. If the exhaust adjustment valve is accidentally ADJUSTING THE touched or if the exhaust controller is replaced, however, the gas flow EXHAUST CONTROL should be che
Page 269B–70115EN/03 9. SETTING AND ADJUSTMENT 9.6 This coefficient is precisely adjusted at the factory and should not require any further adjustment. If the rear mirror of the laser oscillator or the SETTING THE power sensor unit is replaced, however, the power input compensation POWER INPUT coefficient s
Page 2709. SETTING AND ADJUSTMENT B–70115EN/03 9.7 This sensor is precisely adjusted at the factory and should not require any further adjustment. If the water distribution unit or water flow sensor is WATER FLOW replaced, however, this sensor should be set and adjusted as described SENSOR below: 9.7.1 (1)
Page 271B–70115EN/03 9. SETTING AND ADJUSTMENT (10) Gradually reduce the flow rate of the water supply. Check the rate at which the switch trips off. Then, gradually increase the flow rate. If the alarm flow rate is within the following range, the sensor is normal. If the alarm range is exceeded, repeat ste
Page 2729. SETTING AND ADJUSTMENT B–70115EN/03 9.7.2 (1) Adjust the flow rate of cooling water flowing into the oscillator to an Adjusting the Flow alarm setting flow rate. Use a flowmeter with a proper and accurate rated flow rate. Sensor of the C2000C, C3000D, C4000A, and C6000B Model Alarm setting flow r
Page 273B–70115EN/03 9. SETTING AND ADJUSTMENT 9.8 If the laser oscillator is not used for a long time (one week or more) or if the laser gas circulation system is opened to the atmosphere (when a DISCHARGE AGING mirror is cleaned or when a component of the gas system is replaced, for example), aging must b
Page 2749. SETTING AND ADJUSTMENT B–70115EN/03 D Manual operation This method can be used if the machine operator’s panel has a switch for starting internal discharge manually. 1) Enter the desired output and duty cycle on the setting screen. 2) Press the internal discharge switch to start internal discharg
Page 275B–70115EN/03 9. SETTING AND ADJUSTMENT Fig. 9.8 (a) Detail of the Cleaner Flange 2) Mount the cleaner flange on the oscillator. Be sure to keep the flange in the correct orientation. Figs. 9.8 (b) to 9.8 (h) show the locations where the cleaner flange is to be fastened. Fig. 9.8 (b) Inlet of cleaner
Page 2769. SETTING AND ADJUSTMENT B–70115EN/03 Fig. 9.8 (c) Inlet of cleaner flange in C2000B Fig. 9.8 (d) Inlet of cleaner flange in C2000C 274
Page 277B–70115EN/03 9. SETTING AND ADJUSTMENT Fig. 9.8 (e) Inlet of cleaner flange in C3000C Fig. 9.8 (f) Inlet of cleaner flange in C3000D 275
Page 2789. SETTING AND ADJUSTMENT B–70115EN/03 Fig. 9.8 (g) Inlet of cleaner flange in C4000A Fig. 9.8 (h) Inlet of cleaner flange in C6000B 3) Start the oscillator, and check it for leakage. Vacuum it twice. 276
Page 279B–70115EN/03 9. SETTING AND ADJUSTMENT 4) Start the oscillator, and keep the turbo blower rotating (LRDY). Leave the oscillator in this state overnight. 5) More impurities can be captured by using also discharge aging to raise the gas temperature. 6) When gas flow system aging is completed, stop and
Page 281B–70115EN/03 10. REPLACEMENT PROCEDURES 10.1 INPUT UNIT 10.1.1 Replace the stabilized power supply as follows: Replacing the (1) Turn off the power to the CNC, then turn off the power to the laser Stabilized Power oscillator. In addition, open the main circuit breaker for the laser oscillator. Suppl
Page 28210. REPLACEMENT PROCEDURES B–70115EN/03 10.1.4 Replace the FS16–L interface PCBs on the oscillator side as follows: Replacing the FS16–L, When IF PCB (A02B–0128–C202) FS16i–L Interface (1) Replacing the upper PCB (A16B–1110–0260) PCBs on the Oscillator 1) Turn off the power to the CNC, then turn off
Page 283B–70115EN/03 10. REPLACEMENT PROCEDURES A20B–1005–0124 Fig. 10.1.4 (a) Removing the IF PCB unit (FS16–L: A02B–0128–C202) Fig. 10.1.4 (b) Removing the PCB from the PCB holder 281
Page 28410. REPLACEMENT PROCEDURES B–70115EN/03 10.2 Figs. 10.2 (a) to (l) show the laser power supply. Replace the laser power supply as described below. REPLACING THE (1) Replacing the power supply LASER POWER SUPPLY 1) Turn off the power to the laser oscillator, then drain the cooling water. 2) Remove th
Page 285B–70115EN/03 10. REPLACEMENT PROCEDURES 2. Remove the damaged diode by removing the screws and nuts at the both ends. In case of 9 kW power supply at first replace the cooling heat sink and diode plate 1 and 2, then replace the diode block A and B to remove diode. CAUTION When removing a diode, if i
Page 28610. REPLACEMENT PROCEDURES B–70115EN/03 3. Mount a new FET module PCB, fix it using the six screws, then connect the cables. When the 9–kW power supply is being used, connect the CN51 or CN52 cable to the new FET PCB securely, then temporarily tighten all the screws so that the FET PCB is free to mo
Page 287B–70115EN/03 10. REPLACEMENT PROCEDURES Fig. 10.2 (a) Laser power supply unit layout (front side): B204, B205 Fig. 10.2 (b) Laser power supply unit layout (upside): B204, B205 285
Page 28810. REPLACEMENT PROCEDURES B–70115EN/03 Fig. 10.2 (c) Laser power unit layout (inner side): B204, B205 Fig. 10.2 (d) Laser power unit layout (inner part details): B204, B205 286
Page 289B–70115EN/03 10. REPLACEMENT PROCEDURES Fig. 10.2 (e) Diode unit layout (Fig. 10.2 (d) B part details): B204, B205 Fig. 10.2 (f) Diode unit layout (Fig. 10.2 (d) C part details): B204, B205 287
Page 29010. REPLACEMENT PROCEDURES B–70115EN/03 Fig. 10.2 (g) FET module PCB layout (Fig. 10.2 (d) A part details): B204, B205 Cover mounting screw Cover for XT22 Fig. 10.2 (h) Layser PSU layout: B206 to B210 288
Page 291B–70115EN/03 10. REPLACEMENT PROCEDURES Step up transformer PCB Diod unit 2 Diod unit 1 Fig. 10.2 (i) Laser PSU internal details: B206 to B208 289
Page 29210. REPLACEMENT PROCEDURES B–70115EN/03 F type Cathod Anode R type Anode Cathod Fig. 10.2 (j) Mounting position of diode: B206 to B208 290
Page 293B–70115EN/03 10. REPLACEMENT PROCEDURES Cooling heat sink Diode plate 2 Diode plate 1 Diode block A Diode block B Fiber reinforced plastic Diode block face of A Fiber reinforced plastic Diode block face of B Fig. 10.2 (k) Replacing diode: B206 to B208 Fig. 10.2 (l) FET PCB heat sink fixing: B206 to
Page 295B–70115EN/03 10. REPLACEMENT PROCEDURES Fig. 10.2 (n) Mounting position of diode: B209 to B210 293
Page 29610. REPLACEMENT PROCEDURES B–70115EN/03 10.3 Replace the matching box as follows: REPLACING THE (1) Turn off the power to the CNC, then turn off the power to the laser oscillator. In addition, open the main circuit breaker for the laser MATCHING BOX oscillator. (2) Disconnect the power and signal ca
Page 297B–70115EN/03 10. REPLACEMENT PROCEDURES 10.4 When the turbo blower is replaced, the gas circulating system is open to the atmosphere for a long time. Pay particular attention to the immediate REPLACING THE surroundings to ensure that dust and other contaminants do not enter the TURBO BLOWER gas circ
Page 29810. REPLACEMENT PROCEDURES B–70115EN/03 7) Tighten the nut of the damping rubber. Carefully follow the procedure below to prevent the damping rubber from being twisted and damaged. Tightening procedure (C1500B, C2000B, C2000C, C3000C, C6000B) 1. Position the antivibration rubber such that its partin
Page 30010. REPLACEMENT PROCEDURES B–70115EN/03 10.5 Replace the turbo PCB as follows: REPLACING THE (1) Turn off the power to the CNC, then turn off the power to the laser oscillator. In addition, open the main circuit breaker for the laser TURBO PCB oscillator. (2) Disconnect all cables from the turbo PCB
Page 301B–70115EN/03 10. REPLACEMENT PROCEDURES 10.7 When the exhaust pump is replaced, the gas circulating system is open to the atmosphere for a long time. Pay particular attention to the immediate REPLACING THE surroundings to ensure that dust and other contaminants do not enter the EXHAUST PUMP gas circ
Page 303B–70115EN/03 10. REPLACEMENT PROCEDURES 10.8 When the pressure control unit is replaced, the gas circulating system is open to the atmosphere for a long time. Pay particular attention to the REPLACING THE immediate surroundings to ensure that dust and other contaminants do not PRESSURE enter the gas
Page 30410. REPLACEMENT PROCEDURES B–70115EN/03 10.9 When the exhaust control unit is replaced, the gas circulating system is open to the atmosphere for a long time. Pay particular attention to the REPLACING THE immediate surroundings to ensure that dust and other contaminants do not EXHAUST CONTROL enter t
Page 305B–70115EN/03 10. REPLACEMENT PROCEDURES Fig. 10.9 (b) Exhaust control unit C2000B, C2000C, C3000D, C4000A Fig. 10.9 (c) Exhaust control unit C3000C 303
Page 30610. REPLACEMENT PROCEDURES B–70115EN/03 Fig. 10.9 (d) Exhaust control unit C6000B 304
Page 307B–70115EN/03 10. REPLACEMENT PROCEDURES 10.10 When a discharge tube is replaced, the gas circulating system is open to the atmosphere for a long time. Pay particular attention to the immediate REPLACING A surroundings to ensure that dust and other contaminants do not enter the DISCHARGE TUBE gas cir
Page 30810. REPLACEMENT PROCEDURES B–70115EN/03 10.11 The laser oscillator contains many fan units, which are classified into the following two types: REPLACING A FAN (1) Fan unit UNIT Consists of several fans and a terminal block. (2) Fan–assisted radiator Consists of a water–cooled radiator, several fans,
Page 309B–70115EN/03 10. REPLACEMENT PROCEDURES 10.12 The power sensor is a very delicate component. Be careful not to touch or apply force to the black, light–sensitive surface. Also, ensure that no REPLACING THE dust enters the sensor. POWER SENSOR UNIT When removing the power sensor unit from the rear mi
Page 31010. REPLACEMENT PROCEDURES B–70115EN/03 10.13 Replace the shutter section as follows: REPLACING THE SHUTTER SECTION 10.13.1 (1) Turn off the power to the CNC, then turn off the power to the laser Replacing the Shutter oscillator. In addition, open the main circuit breaker for the laser oscillator. U
Page 311B–70115EN/03 10. REPLACEMENT PROCEDURES Fig. 10.13.3 (a) C1500B, C2000B, C2000C, C3000C, C3000D, and C4000A shutter unit layout 309
Page 313B–70115EN/03 10. REPLACEMENT PROCEDURES 10.14 Replace the beam absorber as follows: REPLACING THE (1) Turn off the power to the CNC, then turn off the power to the laser oscillator. In addition, open the main circuit breaker for the laser BEAM ABSORBER oscillator. (2) The beam absorber is water–cool
Page 317B–70115EN/03 10. REPLACEMENT PROCEDURES 10.15 Replace the high–frequency inverter as follows: REPLACING THE (1) Turn off the power to the CNC, then turn off the power to the laser oscillator. In addition, open the main circuit breaker for the laser HIGH–FREQUENCY oscillator. INVERTER (2) Disconnect
Page 31810. REPLACEMENT PROCEDURES B–70115EN/03 10.16 Replace the water distribution unit as follows: REPLACING THE (1) Turn off the power to the CNC, then turn off the power to the laser oscillator. In addition, open the main circuit breaker for the laser WATER oscillator. DISTRIBUTION UNIT (2) Drain all c
Page 319B–70115EN/03 10. REPLACEMENT PROCEDURES Fig. 10.16 (a) Detail of water destributer unit (C1500B) Refer to water flow diagram (Fig. 7.2 (a)). 317
Page 32010. REPLACEMENT PROCEDURES B–70115EN/03 Fig. 10.16 (b) Detail of water destributer unit (C2000B) Refer to water flow diagram (Fig. 7.2 (b)). 318
Page 321B–70115EN/03 10. REPLACEMENT PROCEDURES Fig. 10.16 (c) Detail of water destributer unit (C2000C) Refer to water flow diagram (Fig. 7.2 (c)). 319
Page 32210. REPLACEMENT PROCEDURES B–70115EN/03 Fig. 10.16 (d) Detail of water destributer unit (C3000C) Refer to water flow diagram (Fig. 7.2 (d)). 320
Page 323B–70115EN/03 10. REPLACEMENT PROCEDURES Valve (closed) Valve (open) Position when water is removed Normal position Fig. 10.16 (e) Detail of water destributer unit (C3000D) Refer to water flow diagram (Fig. 7.2 (e)). 321
Page 32410. REPLACEMENT PROCEDURES B–70115EN/03 Valve (closed) Valve (open) Position when water is removed Normal position Fig. 10.16 (f) Detail of water destributer unit (C4000A) Refer to water flow diagram (Fig. 7.2 (f)). 322
Page 325B–70115EN/03 10. REPLACEMENT PROCEDURES Fig. 10.16 (g) Detail of distribution unit A (C6000B) Refer to water flow diagram. 323
Page 32610. REPLACEMENT PROCEDURES B–70115EN/03 Fig. 10.16 (h) Detail of distribution unit B (C6000B) Refer to water flow diagram. 324
Page 327B–70115EN/03 10. REPLACEMENT PROCEDURES 10.17 Replace the flow sensor as follows: REPLACING THE (1) Turn off the power to the CNC, then turn off the power to the laser oscillator. In addition, open the main circuit breaker for the laser FLOW SENSOR oscillator. (2) Drain all cooling water from the os
Page 329B–70115EN/03 10. REPLACEMENT PROCEDURES 10.18 Replace the Dew sensor as follows: REPLACING THE (1) Turn off the power to the CNC, then turn off the power to the laser oscillator. In addition, open the main circuit breaker for the laser DEW SENSOR oscillator. (2) Disconnect the dew sensor Faston term
Page 33010. REPLACEMENT PROCEDURES B–70115EN/03 10.20 Use the procedure below for trigger electrode replacement. REPLACING THE (1) Turn off the power to the CNC, then turn off the main circuit breaker and power supply. TRIGGER (2) Remove bolt and screw , then remove the trigger electrode. ELECTRODE (3) C
Page 33211. LASER OPTICAL SYSTEM B–70115EN/03 11 LASER OPTICAL SYSTEM Table 11 lists the optical parts and their features. Use this table to determine the feature of each optical part. When mounting these parts, be careful to orient them correctly and ensure that they are being mounted on compatible units.
Page 333B–70115EN/03 11. LASER OPTICAL SYSTEM 11.1 The optical parts in the laser oscillator, including the output mirror, rear mirror, folding mirrors, and circular polarization mirror, must be cleaned CLEANING OPTICAL or replaced periodically. Similarly, if the laser output or beam mode PARTS pattern beco
Page 33411. LASER OPTICAL SYSTEM B–70115EN/03 Fig. 11.1 (a) Structure of diagram output coupler (C1500B, C2000B, C2000C) 332
Page 335B–70115EN/03 11. LASER OPTICAL SYSTEM Fig. 11.1 (b) Structure of diagram rear mirror (C1500B, C2000B, C2000C) 333
Page 33611. LASER OPTICAL SYSTEM B–70115EN/03 Fig. 11.1 (c) Structure of diagram output coupler (C3000C, C3000D, C4000A, C6000B) 334
Page 337B–70115EN/03 11. LASER OPTICAL SYSTEM Fig. 11.1 (d) Structure of diagram rear mirror (C3000C, C3000D, C4000A, C6000B) 335
Page 33811. LASER OPTICAL SYSTEM B–70115EN/03 11.1.1 To clean the output coupler, follow the procedure given below. Since it Cleaning and is necessary to touch the optical parts and vacuum parts while cleaning the output coupler, wear clean gloves or finger cots. Do not touch these Replacing the Output part
Page 34011. LASER OPTICAL SYSTEM B–70115EN/03 Fig. 11.1.1 (d) Schematic diagram of mounting of output coupler/ rear mirror to the holder 338
Page 341B–70115EN/03 11. LASER OPTICAL SYSTEM (8) Move the lens cleaning paper horizontally to and fro across the mirror. NOTE Since the lens cleaner has a high surface tension, it may be difficult to move the cleaning paper. As the paper is slowly moved to and fro across the mirror, however, the ethyl alco
Page 34211. LASER OPTICAL SYSTEM B–70115EN/03 11.1.2 To clean the rear mirror, follow the procedure given below. Since it is Cleaning and necessary to touch optical parts and vacuum parts while cleaning the rear mirror, wear clean gloves or finger cots. Do not touch these parts directly Replacing the Rear w
Page 343B–70115EN/03 11. LASER OPTICAL SYSTEM NOTE Since the lens cleaner has a high surface tension, it may be difficult to move the cleaning paper. As the paper is slowly moved to and fro across the mirror, however, the ethyl alcohol will gradually evaporate. (9) Blow clean air (using, for example, a came
Page 34411. LASER OPTICAL SYSTEM B–70115EN/03 11.1.3 To clean the folding mirrors, follow the procedure given below. Since it Cleaning and is necessary to touch the optical parts and vacuum parts while cleaning the folding mirrors, wear clean gloves or finger cots. Do not touch these Replacing the Folding p
Page 345B–70115EN/03 11. LASER OPTICAL SYSTEM (11) When replacing a mirror, remove the mirror from its holder. Since vacuum grease is applied to the rear of the mirror, insert a tool such as a standard screwdriver into the slot on the side plane of the mirror mounting section, and pry out the mirror. Be car
Page 34611. LASER OPTICAL SYSTEM B–70115EN/03 FOLDING MIRROR UNIT FOLDING MIRROR O–RING (G55) MIRROR HOLDER MIRROR HOLDER ATTACHMENT SCREW Fig. 11.1.3 (a) Structure diagram of folding mirror unit 344
Page 347B–70115EN/03 11. LASER OPTICAL SYSTEM FOLDING MIRROR UNIT FOLDING MIRROR O–RING (G65) MIRROR HOLDER MIRROR HOLDER ATTACHMENT SCREW Fig. 11.1.3 (b) Structure diagram of folding mirror unit: C6000B 11.1.4 The zero–shift mirror and circular polarization mirror are used in the optical path length unit,
Page 34811. LASER OPTICAL SYSTEM B–70115EN/03 [Procedure] (1) Turn off the power to the CNC, then open the main circuit breaker or turn off the power supply of the laser oscillator. (2) Remove the mirror holder from the block (alignment unit) by loosening the mirror holder mounting screw. (3) Remove the mir
Page 349B–70115EN/03 11. LASER OPTICAL SYSTEM 11.2 OPTICAL AXIS ADJUSTMENT 11.2.1 The internal surfaces of the output mirror and rear mirror are concave. For Alignment of the laser oscillation, the curvature centers and mirror surfaces of the two mirrors must lie on the same axis. When this axis matches the
Page 35011. LASER OPTICAL SYSTEM B–70115EN/03 4) Obtain a maximum power based on the folding mirror (FM1) by using the folding mirror (FM4) as the reference. 5) Collect a burn pattern. If the mode is good, end this procedure. If not, start all over again from step 1). At this time, collect a burn pattern in
Page 351B–70115EN/03 11. LASER OPTICAL SYSTEM 5) Collect a burn pattern. If the mode is good, end this procedure. If not, start all over again from step 1). At this time, collect a burn pattern in each step. If the mode is finally good, end this procedure. ÌÌ ÑÑ ÌÌ Î FM5 FM4 FM6 ÑÑÓ Î Ó R.M. ÌÌ ÓÓ Ó ÌÌ FM2
Page 35211. LASER OPTICAL SYSTEM B–70115EN/03 5) Obtain a maximum power based on O.C. by using R.M. as the reference. Then, repeat this procedure starting with step 2). At this time, collect a burn pattern in each step. If the mode is finally good, end this procedure. (5) C4000A 1) Obtain a maximum power ba
Page 353B–70115EN/03 11. LASER OPTICAL SYSTEM 11.2.1.4 (1) Minimize the output power of internal discharge. If a large output is Obtaining a maximum used, the mode present when a maximum power is obtained is not always the best mode due to thermal distortion of each mirror. power However, a capability of de
Page 35411. LASER OPTICAL SYSTEM B–70115EN/03 11.2.1.5 (1) For information about burn pattern collection, see Section 4.3(1). Burn pattern collection (2) In the automatic operation mode, execute the program below to emit and beam mode a laser beam to an acrylic block. evaluation [Example of program: For the
Page 355B–70115EN/03 11. LASER OPTICAL SYSTEM Do not determine the maximum power position of A at the actual maximum power point (peak point). Instead, find the first position 1’ with the same power after a peak, and set a position halfway between 1 and 1’. Adjust B to obtain a maximum power at that positio
Page 35611. LASER OPTICAL SYSTEM B–70115EN/03 BEAM PATH (SHORT PATH TIME) ADJUSTMENT SCREW OF ADJUSTMENT SCREW OF ALIGNMENT (N: Y AXIS) ALIGNMENT (N: Y AXIS) REAR MIRROR SIDE (R.M.) OUTPUT COUPLER SIDE (O.C.) ADJUSTMENT SCREW OF ADJUSTMENT SCREW OF ALIGNMENT (M: X AXIS) ALIGNMENT (M: X AXIS) ADJUSTMENT SCRE
Page 357B–70115EN/03 11. LASER OPTICAL SYSTEM BEAM PATH (SHORT PATH TIME) ADJUSTMENT SCREW OF ADJUSTMENT SCREW OF ALIGNMENT (N: Y AXIS) ALIGNMENT (N: Y AXIS) REAR MIRROR SIDE (R.M.) OUTPUT COUPLER SIDE (O.C.) ADJUSTMENT SCREW OF ADJUSTMENT SCREW OF ALIGNMENT (M: X AXIS) ALIGNMENT (M: X AXIS) ADJUSTMENT SCRE
Page 35811. LASER OPTICAL SYSTEM B–70115EN/03 ADJUSTMENT SCREW OF FOLDING MIRROR SIDE (FM3) ALIGNMENT (n: Y AXIS) FOLDING MIRROR SIDE (FM4) OUTPUT COUPLER SIDE (O.C.) ADJUSTMENT SCREW OF ALIGNMENT (M: X AXIS) ADJUSTMENT SCREW OF ALIGNMENT (N: Y AXIS) ADJUSTMENT SCREW OF ALIGNMENT (m: X AXIS) BEAM PATH (SHOR
Page 359B–70115EN/03 11. LASER OPTICAL SYSTEM ADJUSTMENT SCREW OF FOLDING MIRROR SIDE (FM3) ALIGNMENT (n: Y AXIS) FOLDING MIRROR SIDE (FM4) OUTPUT COUPLER SIDE (O.C.) ADJUSTMENT SCREW OF ALIGNMENT (M: X AXIS) ADJUSTMENT SCREW OF ALIGNMENT (N: Y AXIS) ADJUSTMENT SCREW OF ALIGNMENT (m: X AXIS) BEAM PATH (SHOR
Page 36011. LASER OPTICAL SYSTEM B–70115EN/03 ADJUSTMENT SCREW OF ADJUSTMENT SCREW OF ALIGNMENT (N: Y AXIS) ALIGNMENT (m: X AXIS) FOLDING MIRROR SIDE (FM4) ADJUSTMENT SCREW OF ALIGNMENT (n: Y AXIS) REAR MIRROR ADJUSTMENT SCREW OF SIDE (R.M.) ALIGNMENT (M: X AXIS) OUTPUT COUPLER SIDE (O.C.) FOLDING MIRROR SI
Page 361B–70115EN/03 11. LASER OPTICAL SYSTEM ADJUSTMENT SCREW OF FOLDING MIRROR SIDE (FM3) ALIGNMENT (n: Y AXIS) OUTPUT COUPLER FOLDING MIRROR SIDE (FM4) SIDE (O.C.) ADJUSTMENT SCREW OF ALIGNMENT (M: X AXIS) ADJUSTMENT SCREW OF ADJUSTMENT SCREW OF ALIGNMENT (N: Y AXIS) ALIGNMENT (m: X AXIS) ADJUSTMENT SCRE
Page 36211. LASER OPTICAL SYSTEM B–70115EN/03 11.2.2 Alignment of the Laser Diode 11.2.2.1 (1) Attach an acrylic plate approximately 3 to 10 m from the outlet of the Adjusting the beam laser oscillator, and obtain a burn pattern for the CO2 laser beam (one shot, rated output, CW 100%, 0.1–second duration).
Page 363B–70115EN/03 11. LASER OPTICAL SYSTEM Fig. 11.2.2.2 Laser diode unit (A04B–0811–D401) 361
Page 36411. LASER OPTICAL SYSTEM B–70115EN/03 11.2.3 The oscillators of the short optical path types and the C3000C of long Alignment of the Beam optical path type contain a beam folding unit to divert the laser beam within the oscillator. If the beam output direction changes as a result of Folding Unit ali
Page 365B–70115EN/03 11. LASER OPTICAL SYSTEM Fig. 11.2.3 (a) Corner reflectors unit (C1500B, C2000B, C2000C) 363
Page 36611. LASER OPTICAL SYSTEM B–70115EN/03 Fig. 11.2.3 (b) Corner reflectors unit (C3000C, C3000D, C4000A, Short optical path length type) 364
Page 367B–70115EN/03 11. LASER OPTICAL SYSTEM Fig. 11.2.3 (c) Corner reflectors unit (C3000C, Long optical path length type) 365
Page 387B. FANUC LASER C SERIES B–70115EN/03 APPENDIX SPECIFICATIONS C3000C C3000C C3000D C3000D Item (short optical path) (long optical path) (short optical path) (long optical path) Type C3000C C3000D Method High–speed axial carbon dioxide gas laser using high–frequency discharge excitation Oscillator and
Page 388B. FANUC LASER C SERIES SPECIFICATIONS APPENDIX B–70115EN/03 C4000A C4000A Item C6000B (short optical path) (long optical path) Type C4000A C6000B Method High–speed axial carbon dioxide gas laser using high–frequency discharge excitation Structure Combination resonator/power supply type Rated laser
Page 389B–70115EN/03 APPENDIX C. ERROR CODE LIST C ERROR CODE LIST No. Contents Alarm level 16L 0L 4061 661 Anomaly in AD converter 1 ALARM 1 4062 662 Anomaly in AD converter 2 ALARM 1 4063 663 Anomaly in RF/DC power unit ALARM 2 4065 665 Anomaly in shutter action ALARM 2 4066 666 There exists discharge tub
Page 390C. ERROR CODE LIST APPENDIX B–70115EN/03 No. Contents Alarm level 16L 0L 4085 685 Decrease of laser output Warning can be reset 4087 687 Overheat in shutter mirror ALARM 2 4088 688 Decrease of discharge tube voltage ALARM 1 4089 689 Selection signal of assist gas not com- Shutter : close manded (in
Page 391B–70115EN/03 APPENDIX D. PARAMETERS D PARAMETERS This appendix describes the parameters related to sequence control inside the laser oscillator and laser cutting conditions for FS16–L, FS16i–L, and FS0–L. Many parameters are specific to one type of laser oscillator and one machine. So, the user shou
Page 392D. PARAMETERS APPENDIX B–70115EN/03 D.1 Parameter numbers starting from 15000 are used for the FS16–L or FS16i–L. Those starting from 0200 or 3900 are used for the FS0–L. PARAMETERS FOR ENABLING/ DISABLING VARIOUS FUNCTIONS No. Parameter description #7 #6 #5 #4 #3 #2 #1 #0 15000 FLT CLB BMO AGA LGC
Page 393B–70115EN/03 APPENDIX D. PARAMETERS LGC 1 : Two stage gain selection of intra–tube Pressure control valid (usually) 0 : Invalid (When this bit is 1, PRM No. 15245 [207] becomes valid.) When this bit is 1, different control gains are used in the pressure control in discharge start preparation complet
Page 394D. PARAMETERS APPENDIX B–70115EN/03 #7 #6 #5 #4 #3 #2 #1 #0 15002 PSH PCG ADC BMA 3979 C10M BMA GVW BMA 1 : Enables beam output with the shutter closed in the automatic operation mode (for maintenance). 0 : Disables beam output with the shutter closed in the automatic operation mode (standard). PCG
Page 395B–70115EN/03 APPENDIX D. PARAMETERS #7 #6 #5 #4 #3 #2 #1 #0 15004 CWY ECH EDG BMC SPB STC AGC AGC 1 : Controls machining pressure for the set time. 0: Uses a flow pattern. STC 1 : Controls the shutter with an external signal. 0: Controls the shutter with a G code. SPB 1 : Outputs no beam in skip ope
Page 396D. PARAMETERS APPENDIX B–70115EN/03 #7 #6 #5 #4 #3 #2 #1 #0 15006 PCL PIN TRM TRM 1 : Disables the start–point and end–point soft keys for the trace setting screen. 0 : Enables the start–point and end–point soft keys for the trace setting screen. PIN 1 : Specifies a G13 address using inch input. 0 :
Page 397B–70115EN/03 APPENDIX D. PARAMETERS #7 #6 #5 #4 #3 #2 #1 #0 15009 TEM BCG BEM AFZ BS2 BS1 AS2 AS1 This parameter is used for the laser gas mixing function to record and monitor the operating state of the mixer. This parameter is automatically set, so this parameter need not be set manually. AS2,AS1
Page 398D. PARAMETERS APPENDIX B–70115EN/03 D.2 PARAMETERS FOR DISCHARGE TUBE SELECTION #7 #6 #5 #4 #3 #2 #1 #0 15025 PS8 PS7 PS6 PS5 PS4 PS3 PS2 PS1 #7 #6 #5 #4 #3 #2 #1 #0 15026 PS16 PS15 PS14 PS13 PS12 PS11 PS10 PS9 PS1 – 16 Discharge tube selection when half of the discharge tubes are used [How to set]
Page 399B–70115EN/03 APPENDIX D. PARAMETERS D.3 PARAMETERS FOR CONTOURING CONDITIONS 15040 LPWR 0204 LPWR LPWR Power output setting for contour processing (cutting/welding) (Setting data) [Setting range] 0 to 7000 [Unit] WATT [How to set] This parameter can be set in laser setting screen and processing prog
Page 400D. PARAMETERS APPENDIX B–70115EN/03 D.4 PARAMETERS FOR EDGE MACHINING CONDITIONS 15050 EDANG EDANG Edge detection angle A corner is assumed to be present when the angle formed by two blocks is smaller than the specified angle. [Setting range] 0 to 180 [Unit] deg 15051 PCPWR PCPWR Peak piercing power
Page 401B–70115EN/03 APPENDIX D. PARAMETERS [Setting range] 0 to 255 [Unit] 0.01 MPa 15056 EPCAGS EPCAGS Type of piercing assist gas Set a type of assist gas to be used for piercing. [Setting range] 0 to 7 15057 ERTLGH ERTLGH Return distance Set a return distance from the top of a corner to the next block.
Page 402D. PARAMETERS APPENDIX B–70115EN/03 D.5 PARAMETERS FOR PIERCING CONDITIONS 15070 LPWRP 0227 LPWRP LPWRP selection of laser power for piercing (Setting data) [Setting range] 0 to 7000 [Unit] WATT [How to set] This parameter can be set in laser setting screen or processing program. 15071 LHZP 0228 LHZ
Page 403B–70115EN/03 APPENDIX D. PARAMETERS D.6 PARAMETERS FOR HIGH–SPEED PIERCING CONDITIONS 15080 HPPWR HPPWR High–speed piercing power [Setting range] 0 to 7000 [Unit] Watt 15081 HPFRQ HPFRQ Initial high–speed piercing frequency [Setting range] 5 to 2000 [Unit] Hz 15082 STFRQ STFRQ Incremental high–speed
Page 404D. PARAMETERS APPENDIX B–70115EN/03 15086 STPNM STPNM Number of high–speed piercing steps [Setting range] 0 to 32767 15087 HPTIM HPTIM High–speed piercing end time [Setting range] 0 to 9999999 [Unit] msec 404
Page 405B–70115EN/03 APPENDIX D. PARAMETERS D.7 PARAMETERS FOR POWER CONTROL #7 #6 #5 #4 #3 #2 #1 #0 15088 IPC IPC For power control mode specification (models A and B): 0 : The format for specifying S and F in the same block is used. 1 : The format for specifying G63 or entering the external signal (G224#1
Page 406D. PARAMETERS APPENDIX B–70115EN/03 15093 LDTY0 0244 LDTY0 LDTY0 Power control constant When power control has been enabled with bit 0 of PRM No. 15000 [200], power control is performed according to the value of LDTY0. LDTY0 is used to specify the duty ratio when the feedrate is 0, thus enabling mod
Page 407B–70115EN/03 APPENDIX D. PARAMETERS #7 #6 #5 #4 #3 #2 #1 #0 15096 PCD PCF PCP PCP 0 : Does not exercise feedrate–based output control in newly specified power control. 1 : Exercises feedrate–based output control in newly specified power control. PCF 0 : Does not exercise feedrate–based frequency con
Page 408D. PARAMETERS APPENDIX B–70115EN/03 D.8 PARAMETERS FOR ASSIST GAS PRESSURE AND TIME SETTING 15100 LAGASL 0214 LAGASL LAGASL Selection of assist gas (Setting data) [Setting range] 0 to 7 [How to set] This parameter can be set either in laser setting screen or in material processing program. 15101 LAF
Page 409B–70115EN/03 APPENDIX D. PARAMETERS 15114 LAGPR1 0215 LAGPR1 15115 LAGPR2 0216 LAGPR2 15116 LAGPR3 0217 LAGPR3 LAGPR1, 2, 3 Pre–flow pressure 1, 2, 3 (Setting data) [Setting range] 0 to 255 [Unit] 0.01 MPa [How to set] This parameter can be set in laser setting screen. 15120 LAGEX1 0218 LAGEX1 15121
Page 410D. PARAMETERS APPENDIX B–70115EN/03 15132 AGPST 3990 AGPST AGPST Maximum assist gas pressure Set an assist gas pressure (MPa) at which 12.5 V is output. [Setting range] 0 to 255 [Unit] MPa [Standard setting] 255 15135 AGTIM AGTIM Assist gas pressure set time Set a desired assist gas pressure set tim
Page 411B–70115EN/03 APPENDIX D. PARAMETERS D.9 PARAMETERS FOR LASER MAINTENANCE TIMING INDICATION FUNCTIONS 15150 LTTIME 0187 LTTIME LTTIME Laser RUN ON time A cumulative RUN ON time is automatically set. [Setting range] 0 to 99999999 [Unit] 0.1 hour Standard setting: Automatically set 15151 VPTIME VPTIME
Page 412D. PARAMETERS APPENDIX B–70115EN/03 D.10 PARAMETERS FOR THE OSCILLATOR 15200 LPCMD1 LPCMD1 Power used for power calibration factor determination when half of the discharge tubes are used Set a power to find the power calibration factor when half of the discharge tubes are used. [Setting range] 0 to
Page 413B–70115EN/03 APPENDIX D. PARAMETERS 0232 LPCMD3 LPCMD3 Power command for power calibration in mode of 8 discharge tubes. This is power command for power calibration in mode of 8 discharge tubes. Duty is 100%. [Setting range] 0 to 32767 [Unit] WATT [Standard setting] Refer to the parameter data sheet
Page 414D. PARAMETERS APPENDIX B–70115EN/03 15207 PCLMT 3986 PCLMT PCLMT Maximum specifiable power If a specified laser power after power compensation and power feedback processing exceeds the setting in this parameter, the laser power is clamped to the setting. [Setting range] 0 to 32767 [Unit] WATT [Stand
Page 415B–70115EN/03 APPENDIX D. PARAMETERS [Unit] WATT [Standard setting] Refer to the parameter data sheet delivered with the machine. 15212 PPCLMT PPCLMT Maximum specifiable power when the duty ratio is not clamped If PRM No. 15213 does not clamp the duty ratio to 50%, and a specified power after power c
Page 416D. PARAMETERS APPENDIX B–70115EN/03 D.11 PARAMETERS FOR DISCHARGE 15220 LVICMD 0247 LVICMD LVICMD Maximum bias command at discharge start Discharge current command is given when discharge start signal (HVON) is commanded in discharge preparation completion (LRDY) state to start discharge. This comma
Page 417B–70115EN/03 APPENDIX D. PARAMETERS 15223 LVCMD 0250 LVCMD LVCMD Bias command in oscillator preparation completion state After the start of discharge is confirmed, the bias voltage of discharge current command is replaced by this parameter. This replacement is performed during a rise in the gas pres
Page 418D. PARAMETERS APPENDIX B–70115EN/03 D.12 PARAMETERS FOR GAS CONTROL (1) 15240 LEXPS 0235 LEXPS LEXPS Negative pressure in exhaust completion This is the ultimate vacuu when gas flow system is evacuated in laser [Starting range] 0 to 32767 [Unit] 13 Pa [Standard setting] Refer to the parameter data s
Page 419B–70115EN/03 APPENDIX D. PARAMETERS 0236 LPPS LPPS Intra–tube pressure during oscillation This is intra–tube gas pressure in laser preparation completion (LSTR) state. [Unit] 13 Pa [Standard setting] Refer to the parameter data sheet delivered with the machine. 15244 LCCGN 0264 LCCGN LCCGN Gas press
Page 420D. PARAMETERS APPENDIX B–70115EN/03 15248 GPNRAV 4399 GPNRAV GPNRAV Allowable gas pressure insufficiency The insufficient gas pressure alarm is issued when the gas pressure in the discharge tube does not reach [(specified tube pressure for oscillation) – (this allowable value)]. [Setting range] 0 to
Page 421B–70115EN/03 APPENDIX D. PARAMETERS 15257 VIBCNT VIBCNT Number of abnormal vibrations detected If the number of abnormal vibrations detected exceeds the value specified in this parameter, an abnormal vibration alarm is issued. [Setting range] 0 to 32767 [Unit] Number [Standard setting] 30 15258 LGPT
Page 422D. PARAMETERS APPENDIX B–70115EN/03 D.13 PARAMETERS FOR HIGHLY REFLECTIVE MATERIAL ALARMS 15265 PWRUP 3982 PWRUP PWRUP Maximum allowable power increase If the actual power exceeds the specified power by the value specified in this parameter, a beam reflection error is reported. [Setting range] 0 to
Page 423B–70115EN/03 APPENDIX D. PARAMETERS D.14 PARAMETERS FOR LASER POWER/VOLTAGE DROP 15270 NRMLTV 0199 NRMLTV NRMLTV Discharge tube voltage in normal operation In this parameter, the voltage to be applied to the discharge tube in normal base discharge is set. The value set in this parameter serves as th
Page 424D. PARAMETERS APPENDIX B–70115EN/03 15276 GCHTM 3992 GCHTM GCHTM Laser gas replacement time Set a time for laser gas replacement in laser start operation. [Setting range] 0 to 32767 [Unit] sec [Standard setting] Refer to the parameter data sheet delivered with the machine. 424
Page 425B–70115EN/03 APPENDIX D. PARAMETERS D.15 PARAMETERS FOR POWER TABLE SETTING 15280 LPCPW1 15281 LPCPW2 LPCPW1 Table interval when half of the discharge tubes are used Set a power command setting interval. [Setting range] 30 to 1200 [Standard setting] Refer to the parameter data sheet delivered with t
Page 426D. PARAMETERS APPENDIX B–70115EN/03 15295 LPCP15 0282 LPCP25 15296 LPCP16 0283 LPCP26 15297 LPCP17 0284 LPCP27 15298 LPCP18 0285 LPCP28 LPCP20–28 Power command in mode of 4 discharge tubes Power command is increased by increment of PRM No. 15280[266] for mode of 4 discharge tubes starting from 0. [S
Page 427B–70115EN/03 APPENDIX D. PARAMETERS D.16 AUTOMATIC AGING FUNCTION 15320 LPCD LPCD2C Command power for power calibration coefficient calculation (valid when EGE = 1) If the automatic aging function is enabled, a power calibration coefficient is calculated after aging ends. [Setting range] 0 to 7000 [
Page 428D. PARAMETERS APPENDIX B–70115EN/03 15324 LPSC LPSC Intra–tube pressure setting at oscillation time for power calibration coefficient calculation (50 Hz) (valid when EGE = 1) For an oscillator with an inverter mounted, the same value as parameter No. 15325 is to be set. This parameter specifies a va
Page 429B–70115EN/03 APPENDIX D. PARAMETERS 15328 LDCD2E LDCD2E Pulse duty cycle command for aging (valid when EGE = 1) This parameter specifies a duty cycle to be used when aging is performed with the automatic aging function. [Setting range] 0 to 100 [Unit] % [Standard setting] 50 15329 PLCTME PLCTME Powe
Page 430D. PARAMETERS APPENDIX B–70115EN/03 15332 LFCMD2 LFCMD2 Oscillation frequency command value for power calibration [Setting range] 5 to 2000 [Unit] Hz [Standard setting] 100 15333 LDCMD2 LDCMD2 Pulse duty cycle command during power calibration and aging [Setting range] 0 to 100 [Unit] % [Standard set
Page 431B–70115EN/03 APPENDIX D. PARAMETERS 15338 LOFTIMH LOFTIMH Aging time data 4 (valid when EGE = 1) When the automatic aging function is enabled, the hour when the oscillator was stopped previously is recorded. [Setting range] 0 to 24 (Automatically set by the CNC) 431
Page 432D. PARAMETERS APPENDIX B–70115EN/03 D.17 PARAMETERS FOR TRACING (DETECTION) 15500 KE 0252 KE KE Detector gain factor This parameter sets a gain factor for the displacement detector in Z–axis tracing control. [Setting range] 2048 (0.5 212) to 6144 (1.5 212) [Standard setting] 4096 NOTE This parameter
Page 433B–70115EN/03 APPENDIX D. PARAMETERS 15504 EMUL 0295 EMUL EMUL Displacement multiplication factor Set a displacement multiplication factor to match a greater value to be detected in tracing control. (Usually, 1 is set.) [Setting] 1 or 5 [Standard setting] 1 Set value Maximum displacement Resolution (
Page 434D. PARAMETERS APPENDIX B–70115EN/03 D.18 PARAMETERS FOR TRACING (DISPLAY) #7 #6 #5 #4 #3 #2 #1 #0 15510 TRA TRA 1 : Uses an absolute coordinate system to display the current position on the tracer setting screen. 0 : Uses a relative coordinate system to display the current position on the tracer set
Page 435B–70115EN/03 APPENDIX D. PARAMETERS D.19 PARAMETERS FOR TRACING (INTEGRAL PROCESSING) 15510 INTGT 0255 INTGT INTGT Integral time constant Set an integral time constant for integral processing of deviations in Z–axis tracing control. [Setting range] 0 to 32767 [Unit] msec [Standard setting] 0 15511 Z
Page 436D. PARAMETERS APPENDIX B–70115EN/03 D.20 PARAMETERS FOR TRACING (PHASE COMPENSATION) 15520 ISOUT 0257 ISOUT ISOUT Phase compensation time constant Set a time constant for displacement phase compensation in Z–axis tracing control. [Setting range] 0 to 32767 [Unit] msec [Standard setting] 0 (Values of
Page 437B–70115EN/03 APPENDIX D. PARAMETERS D.21 PARAMETERS FOR TRACING (DISPLACEMENT) 15530 NMDEF 0224 NMDEF NMDEF Reference displacement Set a reference distance to be controlled from the quill top to a workpiece to be controlled in Z–axis tracing control. [Setting range] PRM No. 15531 [260] to PRM No. 15
Page 438D. PARAMETERS APPENDIX B–70115EN/03 [Setting range] 0 to 32767 [Unit] 0.001 mm [Standard setting] 1200 15537 EOVRS 3987 EOVRS EOVRS Detection displacement for issuing the tracing range exceeded alarm In Z–axis tracing control, when the absolute value, DE, of a tracing deviation (detection displaceme
Page 439B–70115EN/03 APPENDIX D. PARAMETERS D.22 PARAMETERS FOR TRACING (FEEDRATE COMMAND) 15540 ZTCGM 0296 ZTCGM 15541 WTCGM 3984 WTCGM 15542 XTCGM 15543 YTCGM ZTCGM Conversion factor for feedrate control voltage in Z–axis tracing control WTCGM Conversion factor used to convert a voltage for feedrate speci
Page 440D. PARAMETERS APPENDIX B–70115EN/03 D.23 PARAMETERS FOR TRACING (APPROACH) 15550 APRCH 0251 APRCH APRCH Approach completion detection range Set a range of displacements assumed to represent approach completion in Z–axis tracing control. Approach completion is assumed when the state where displacemen
Page 441B–70115EN/03 APPENDIX D. PARAMETERS D.24 PARAMETERS FOR CONSTANT OPTICAL PATH LENGTH CONTROL #7 #6 #5 #4 #3 #2 #1 #0 15700 8 7 6 5 4 Z Y X X – 8 Set axes required to calculate a move command for a mirror movement axis. 1 : Used for move command calculation 0 : Not used for move command calculation N
Page 442D. PARAMETERS APPENDIX B–70115EN/03 15704 MIRCOM MIRCOM Mirror block axis reference position compensation When correcting the mirror block axis position to obtain an optimum propagation distance, set the distance from the origin in this parameter. [Setting range] "32767 [Unit] mm [Standard setting]
Page 443B–70115EN/03 APPENDIX D. PARAMETERS D.25 LASER GAS MIXER FUNCTION 15710 BMBCTM 4390 BMBCTM BMBCTM External piping exhaust time [Setting range] 0 to 32767 [Unit] sec [Standard setting] 10 (2/m pipe length) 15711 GMMWT 4411 GMMWT GMMWT Gas mixture wait time [Setting range] 0 to 32767 [Unit] min [Stand
Page 444D. PARAMETERS APPENDIX B–70115EN/03 15715 VCCKPR VCCKPR Vacuum criteria gas pressure [Setting range] 0 to 32767 [Unit] 13 Pa [Standard setting] 100 15716 TVTMR 4391 TVTMR TVTMR Evacuation wait time [Setting range] 0 to 32767 [Unit] min [Standard setting] Refer to the parameter data sheet delivered w
Page 445B–70115EN/03 APPENDIX D. PARAMETERS D.26 PARAMETERS FOR GAS CONTROL (2) #7 #6 #5 #4 #3 #2 #1 #0 15800 TPC TPC 0 : Disables the turbo blower constant power drive control function. 1 : Enables the turbo blower constant power drive control function. 15801 LPPSM LPPSM Maximum intra–tube pressure when th
Page 446D. PARAMETERS APPENDIX B–70115EN/03 15804 TPW TPW Maximum open time of the exhaust valve for intra–tube pressure control at oscillation time [Unit] msec [Setting range] 0 to 3000 446
Page 447E. CONTROL SEQUENCES B–70115EN/03 APPENDIX IN LASER OSCILLATOR E CONTROL SEQUENCES IN LASER OSCILLATOR 447
Page 448E. CONTROL SEQUENCES IN LASER OSCILLATOR APPENDIX B–70115EN/03 E.1 OUTLINE OF LASER OSCILLATION SEQ 0 SEQUENCES PURGE =0 LRDY =0 PTLP =0 LSTR =0 POWER ON POWER OFF SEQ 10 PURGE =1 LRDY =0 INITIAL PTLP =0 LSTR =0 SEQ 11 RUN ON (WAIT = 1) SEQ 19 RUN OFF (WAIT = 1) SEQ 20 PURGE =0 READY OF DISCHARGING
Page 449E. CONTROL SEQUENCES B–70115EN/03 APPENDIX IN LASER OSCILLATOR [SEQ 0] POWER OFF state This is the state prior to power on where all the signals are in off state. [SEQ 10] INITIAL state After power is turned on, laser comes into PURGE=1 state. This signal is on when atmospheric pressure sensor monit
Page 450E. CONTROL SEQUENCES IN LASER OSCILLATOR APPENDIX B–70115EN/03 E.2 Intra–tube gas pressure control is performed in the following steps as shown in the figure. INTRA–TUBE GAS PRESSURE CONTROL SEQUENCES GAS PRESSURE : COMMAND : REAL PRM15249 PRM15242 15243 [236] Pressure value "20 (="266 Pa) PRM15241
Page 451E. CONTROL SEQUENCES B–70115EN/03 APPENDIX IN LASER OSCILLATOR [SQ 10] This is the state after power is turned on in which intra–tube gas pressure equals atmospheric pressure. [SQ16, 17, 18] Setting the laser start switch (RUN) to ON starts evacuation, which continues until the pressure designated b
Page 452E. CONTROL SEQUENCES IN LASER OSCILLATOR APPENDIX B–70115EN/03 [SQ49, 13, 12] In the SQ20 state, setting the laser start switch (RUN key) to OFF terminates gas pressure control, after which the turbo blower is shut down, taking about 70 seconds to stop. If evacuation after RUN OFF has been enabled,
Page 453E. CONTROL SEQUENCES B–70115EN/03 APPENDIX IN LASER OSCILLATOR E.3 The control of tube voltage is done simultaneously with that of gas pressure as shown in Fig. E.3 in the following sequences. TUBE VOLTAGE CONTROL SEQUENCES GAS PRESSURE PRM15242, 15243 (PRM236) PRM15241 (PRM237) SQ20 SQ26 SQ27 SQ28
Page 454E. CONTROL SEQUENCES IN LASER OSCILLATOR APPENDIX B–70115EN/03 [SQ 291] When HV is turned on, high voltage is applied to discharge tubes and voltage control begins. Here command is increased in ramp mode up to the value of PRM 15220 [247] (maximum bias command at discharge start). The time required
Page 455E. CONTROL SEQUENCES B–70115EN/03 APPENDIX IN LASER OSCILLATOR [SQ 294] When normal base discharge is confirmed and PRM15000 bit 4 [200 bit 4] is 1 (power calibration is done), this sequence begins to obtain power calibration coefficient. Power calibration is to calibrate so that real and command po
Page 456E. CONTROL SEQUENCES IN LASER OSCILLATOR APPENDIX B–70115EN/03 E.4 OSCILLATION Power ON sequence SEQUENCES FLOW CHART PURGE = ON OFI = OFF NO RUN = ON YES Alarm 1 monitor treatment start PCR = OFF OFI = ON PURGE = OFF WAIT = ON CLON = ON Wait for 3 seconds RPA = ON CAMCON = FULL OPEN Chiller monitor
Page 457E. CONTROL SEQUENCES B–70115EN/03 APPENDIX IN LASER OSCILLATOR TIMER A RESET This timer begins counting after the start of evacuation. 1 VEN = ON PCL = OFF Wait for 5 seconds GRDY = ON GRDY = OFF CAMCON = FULL CLOSE NO Timer count Intra–tube pressure PRM15259 [600 se
Page 458E. CONTROL SEQUENCES IN LASER OSCILLATOR APPENDIX B–70115EN/03 VEN=OFF Exhaust valve is closed. RBA=ON Start of Roots blower. 2 Gas pressure control start 200 VAC to DC power supply unit PRM15240 [235] → PRM15241 [237] Wait for 45 seconds PRM15241 [237] +20 > Intra–tube pressure NO > PRM15241 [237]
Page 459E. CONTROL SEQUENCES B–70115EN/03 APPENDIX IN LASER OSCILLATOR HV=ON When HV is turned on, laser can proceed to the next sequence. HV ON can either be done in manual mode or automatic mode using PMC. LRDY=OFF When HV is turned on, LRDY becomes off. PSS1–8=SET This is to select power units and usuall
Page 460E. CONTROL SEQUENCES IN LASER OSCILLATOR APPENDIX B–70115EN/03 discharge is confirmed, BIAS COMMAND is changed automatically from PRM15220[247] to PRM15223 [250] (BIAS COMMAND). If all the tubes are confirmed normal discharge, intra–tube pressure is changed from PRM15241 [237] to PRM15242, 15243 [23
Page 461E. CONTROL SEQUENCES B–70115EN/03 APPENDIX IN LASER OSCILLATOR The result is compared with PRM15203 [240] (power calibration limit). When PRM15204 [241] exceeds PRM15203 [240], the former is clamped by the latter. LSTR=ON Oscillation preparation completion signal is on. Here oscillation in both manu
Page 462F. REFIXING AND REPLACING GAS TUBE APPENDIX B–70115EN/03 F REFIXING AND REPLACING GAS TUBE Replacing both of polyethylene and copper tubes is the same. Take care only for the combination between tube materials and ferrule materials. Polyethylene tube : nylon ferrule (back, front) (A98L–0004–0348/2–N
Page 463F. REFIXING AND REPLACING B–70115EN/03 APPENDIX GAS TUBE F.1 1 Loosen tube fixing nut. Make sure to hold the fitting base with wrench. REFIXING TUBE 2 Remove tube. Removed tube is accompanied by ferrule as shown in Fig. F.1. Make sure that neither of tube nor ferrule is contaminated with dust. When
Page 464F. REFIXING AND REPLACING GAS TUBE APPENDIX B–70115EN/03 F.2 1 Cut the tube perpendicular to its length, then mount new ferrules on the tube. REPLACING TUBE Referring to Fig. F.1, mount fixing nut, back ferrule, and front ferrule onto tube. Here make sure not to introduce dust into tube and ferrule.
Page 465G. REFIXING AND REPLACING B–70115EN/03 APPENDIX WATER TUBE G REFIXING AND REPLACING WATER TUBE For brass and stainless steel tubes, be careful to make proper selection of fitting material and structure. Here make sure not to wet inside of apparatus in draining and removing tube. 465
Page 466G. REFIXING AND REPLACING WATER TUBE APPENDIX B–70115EN/03 G.1 1) Loosen tube fixing nut. Make sure to hold the fitting base with wrench. REFIXING TUBE TO 2) Remove tube. BRASS FITTING a) In case of brass fitting: Removed tube is accompanied as shown in Fig. G.1 by plastic sleeve. Nut Tube Plastic s
Page 467G. REFIXING AND REPLACING B–70115EN/03 APPENDIX WATER TUBE G.2 1) (a) In case of brass fitting Mount a new sleeve on the tube. REPLACING TUBE (b) In case of stainless steel fitting If a sleeve is necessary, mount a new sleeve on the tube. If no sleeve is necessary, insert the tube as is. In mounting
Page 468H. GLOSSARY APPENDIX B–70115EN/03 H GLOSSARY Name Meaning Access panel That protective component of a housing or enclosure which, when removed or shifted, can cause exposure to laser radiation AEL Accessible Maximum accessible emission level set up for each emission level class of laser products Ali
Page 469B–70115EN/03 APPENDIX H. GLOSSARY Name Meaning Feedback Corrective action in which a quantity to be controlled is compared with a target value by feeding part of an output signal to the input through a specially created closed loop so that the quantity to be controlled matches the target value Laser
Page 470H. GLOSSARY APPENDIX B–70115EN/03 Name Meaning Protective housing Laser product housing or its part designed to protect humans from exposure to a laser beam exceeding an accessible emission level or strong collateral radiation Pulse duration Time interval between the points at which an instan- taneo
Page 471B–70115EN/03 Index ƠAơ Connection of cooling water pipes, 108 Control sequences in laser oscillator, 441 Adjusting the beam center of the laser diode beam superimposed on the CO2 laser beam, 354 Cooling water, 87 Adjusting the diameter of the laser diode beam, 354 Cooling water piping, 194 Adjusting
Page 472INDEX B–70115EN/03 ƠIơ Obtaining a maximum power, 345 Optical axis adjustment, 341 Indication of state by means of self diagnostic function, 171 Organization of the manual, 2 Input unit, 211, 273 Oscillation sequences flow chart, 450 Installation, 66 Oscillator connections, 87, 186 Installation proc
Page 473B–70115EN/03 INDEX Refixing tube to brass fitting, 460 Responding to alarm messages on the screen, 136 Related manuals, 4 Removing cooling water, 77 Replacement procedures, 272 ƠSơ Replacing a discharge tube, 299 Safety enclosure (at your work station), 20 Replacing a fan–assisted radiator, 300 Sett
Page 475Revision Record FANUCĄLASER C series MAINTENANCE MANUAL (B–70115EN) D Additional descriptions of C2000C, C3000D, C600B and 03 Apr., 2000 C4000A. D Addition of C3000C. 02 Jul., 1996 D Correction of errors 01 May, 1995 Edition Date Contents Edition Date Contents
Page 477EUROPEAN HEADQUARTERS – BELGIUM / NETHERLANDS GRAND-DUCHÉ DE LUXEMBOURG GE Fanuc Automation Europe S.A. GE Fanuc Automation Europe S.A. - Netherlands Branch - Zone Industrielle Postbus 7230 - NL-4800 GE Breda L-6468 Echternach Minervum 1603A - NL-4817 ZL Breda ( (+352) 727979 - 1 ( (+31) 76-5783 201
Page 478• No part of this manual may be reproduced in any form. • All specifications and designs are subject to change without prior notice. The export of this product is subject to the authorization of the government of the country from where the product is exported. In this manual we have tried as much as
Page 479TECHNICAL REPORT NO.TML 01/005E Date. JUN. 11.2001 General Manager of Laser Laboratory Correction of the manual of FANUC LASER C series 1. Communicate this report to: ○ Your information ○ GE Fanuc-N, GE Fanuc-E FANUC Robotics CINCINNATI MILACRON ○ Machine tool builder Sales agency End user 2. Summar
Page 480Corrections of the periodic maintenance 1.Type of applied technical documents Name FANUC LASER C series OPERATOR'S MANUAL Spec. No. / Ver. B-70114EN/04 2.Summary of change Group Name / Outline New, Add Applicable Correct, Del Date Basic Function Optional Function Unit Maintenance Parts Notice Correc
Page 481In FANUC LASER C series OPERATOR’S MANUAL B-70114EN/04, interval of turbo blower oil change is corrected on page 121,122. Table 5.2(a) Periodic maintenance items and periods (Page 121) Interval of maintenance (Operation hour) C2000B C3000C C1500B C2000C C3000D True : 1000h or inferior turbo blower o
Page 482Corrections of the periodic maintenance 1.Type of applied technical documents Name FANUC LASER C series MAINTENANCE MANUAL Spec. No. / Ver. B-70115EN/03 2.Summary of change Group Name / Outline New, Add Applicable Correct, Del Date Basic Function Optional Function Unit Maintenance Parts Notice Corre
Page 483In FANUC LASER C series MAINTENANCE MANUAL B-70115EN/03, the interval of turbo blower oil change is corrected on page 112,113. Table 5.2(a) Periodic maintenance items and periods (Page 112) Interval of maintenance (Operation hour) C2000B C3000C C1500B C2000C C3000D True : 1000h or inferior turbo blo