B–70115EN/03
9. SETTING AND ADJUSTMENT
241
7) C6000B
Output compensation disable
Bit 4 of parameter No. 15000, parameter No. 15201 (bit 4 of
parameter No. 200, parameter No. 231, No. 232): 0
Laser power feedback disable
Parameter No. 15208 (parameter No. 233): 0
Maximum output command setting
Parameter No. 15207, No. 15210, No. 15212: Value of
parameter No. 15307
(Parameter No. 3986, No. 238): Value of parameter No. 293
(4)For power supplies 2, 3, 6, and 7 on the C6000B (A14B–0082–B205
or B208), change the setting of the setting pin S2 from B202 to B201,
and rotate VR11 one turn counterclockwise. (See Note 5.)
(1)Start the oscillator, and wait until the gas pressure is stabilized in the
base discharge state. (Wait about 3 minutes. See DGN 905 [16–L,
16i–L] or DGN 860 [0–L].) Then, check that all discharge tubes are
continuously discharging. If a discharge tube stops discharge, turn
VR11 of the power supply corresponding to the discharge tube
clockwise until the discharge tube discharges. Adjust VR11 so that all
discharge tubes are continuously discharging in the state immediately
before blinking starts.
(2)Record the value of RFI or the values of RFI and DCV. (See Note 1.)
(a). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(3)Turn VR11 of power supply 1 gradually counterclockwise to decrease
the value of RFI by 4 mA or the value of DCV by 0.7 V. Perform this
operation until the discharge tube stops discharge. When the discharge
tube stops discharge, turn VR11 of power supply 1 gradually
clockwise to increase the value of RFI by 4 mA or the value of DCV
by 0.7 V. Adjust VR11 of power supply 2 in the same way as for power
supply 1.
(1)Start the oscillator, and wait until the gas pressure is stabilized in the
base discharge state. (Wait about 3 minutes. See DGN 905 [16–L,
16i–L] or DGN 860 [0–L].) Check that all discharge tubes are
continuously discharging. If a discharge tube stops discharge, turn
VR11 of the power supply corresponding to the discharge tube
clockwise until the discharge tube discharges.
(2)Adjust VR11 of power supply 1 so that the discharge tube is
continuously discharging in the state immediately before blinking
starts. Adjust VR11 of each of power supplies 2, 3, and 4 so that each
discharge tube blinks.
(3)Record the value of RFI or the values of RFI and DCV. (See Note 1.)
(a). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9.1.2
Base Discharge
Adjustment
9.1.2.1
C1500B (without a
trigger electrode)
(step 1)
9.1.2.2
C2000B (step 1)