9.SETTING AND ADJUSTMENT B-70255EN/01
- 118 -
turning off HV, turn on HV and check that stable discharge is
maintained for 5 minutes. Until this stable discharge state is
confirmed, repeat this step.
(8) Turn off HV, then set 2850 in PRM 15223 for specifying a bias
command value. (See Section 9.1.6.)
(9) When the base discharge state is set, wait for the gas pressure to
become stable.
(10) Record the value of RFI.----------------------------------------------(A)
9.1.3 Maximum Output Adjustment
(1) Specify the output of a half of the rated output value, and turn on
the beam. While checking that the power supply output (Note 1
in the Table 9.1.7) does not exceed the maximum power supply
output (Table 9.1.7) and that the value of DCV (Note 1 in the
Table 9.1.7) does not exceed 230V with A14B-0082-B209,
increase the output gradually to the maximum command value
(Table 9.1.7). If the maximum value is about to be exceeded,
turn VR13 counterclockwise to prevent the maximum value from
being exceeded.
(2) After output of the maximum command value, wait 3 minutes.
Then, adjust VR13 to set the power supply output to the
maximum power supply output.
(3) Record the value of RFI or the values of RFI and DCV
(See Subsection 9.1.5)-------(B)
9.1.4 Check
9.1.4.1 Fine adjustment of VR11 and VR13
(1) Three minutes after the beam is turned off, adjust RFI or DCV to
the value of (A) with VR11.
(2) Three minutes after the maximum command value is output,
adjust RFI or DCV to the value of (B) with VR13.
(3) Repeat (1) and (2) until no adjustment is required.
(4) Turn off the beam, and wait 3 minutes. Record the DCV, DCI,
RFV, and RFI values of each power supply. Then, check that the
laser output Pa does not exceed the maximum base output (Table
9.1.7), and that heat-sensitive paper does not react in the test of
Subsection 9.1.7. If a problem is observed, make another power
supply adjustment.