B-70255EN/01 APPENDIX E.CONTROL SEQUENCES IN LASER OSCILLATOR
- 251 -
E.4 OSCILLATION SEQUENCES FLOW CHART
POWER ON SEQUENCE
PURGE = ON
OF1 = OFF
RUN = ON
Alarm 1
Monitor treatment
start
PCR = OFF
OF1 =ON
PURGE = OFF
WAIT = ON
CLON = ON
Wait for 3 seconds
RPA = ON
Control valve = full open
Chiller monitor start
Timer A = Reset
Wait for 3 seconds
1
Yes
No
RURGE = ON : This signal becomes on after POW ON.
OF1 = OFF : Off-interlock contact OFF state
RUN = ON : Laser start switch on.
PCR = OFF : Power supply unit alarm indication and holding
circuit is cleared.
OFI = ON : CNC POWER OFF is invalid.
WAIT = ON : LRDY waiting.
CLON = ON : Chiller start requested.
RPA = ON : Exhaust pump start.
Chiller monitor : Monitoring of the start signal from the chiller
TIMER A=RESET : This timer begins counting after the start of
evacuation.