E.CONTROL SEQUENCES IN LASER OSCILLATOR APPENDIX B-70255EN/01
- 254 -
HV = ON :
When HV is turned on, you proceed to the sequence below.
LRDY = OFF :
When HV is turned on, LRDY is turned off.
PSS1-8 = SET :
This is to select power units and usually is sent to the units
selected by PRM 15025 to PRM 15028.
PSRDY = ON :
This is sent to a power supply selected by the laser power supply
ready signal.
RFHV = ON :
Power is fed from the DC section of the laser power supply to
the RF section.
BIAS COMMAND
0
®
PRM.15220
PRM.15220 WAIT
Intra-tube pressure
< PRM.15221
GAS PRESSURE CONTROL
PRM.15241
®
PRM .15242,15243
PRM.15242,15243-100
<PRM.15242,15243
<PRM.15242,15243+100
ALM.4078
BIAS COMMAND
PRM.15220
®
PRM.15223
ALM.4066
3
4
Yes
Yes
No
No
The output command value is changed from a low level to the value
set in PRM 15220 (maximum bias command value at discharge time)
in a ramp mode by the bias command to facilitate discharge. The time
required to reach peak value is 6 sec.