B-70255EN/01 APPENDIX D.PARAMTER LIST
- 231 -
15246 Pressure control integration gain
Data type: Word
Unit of data:
Range of data: 0 to 32767
This is feedback integration gain for gas pressure control.
15247 Pressure control integral element preset value
Data type: Word
Unit of data:
Range of data: -32768 to 32767
Integral element preset value for gas pressure control
15248 Allowable gas pressure insufficiency
Data type: Word
Unit of data: (1=13Pa)
Range of data: 0 to 32767
Allowable value used to determine to issue no alarm if the gas
pressure does not reach the setting
15249 Gas pressure rising time
Data type: Word
Unit of data: sec
Range of data: 0 to 60
Set the gas pressure rising time in sequence 27.
15255 RUN ON evacuation time
Data type: Word
Unit of data: sec
Range of data: 0 to 32767
Set the evacuation time at the RUN start.
15256 RUN OFF evacuation time
Data type: Word
Unit of data: sec
Range of data: 600 to 32767
When bit 2 of parameter No. 15003 (GVW) is set 1 and the oscillator
start signal (G222#6 RUN) is set to "0," evacuation is performed
according to this time setting.